| Literature DB >> 9894346 |
Q C Chen1, S F Mou, X P Hou, J M Riviello, Z M Ni.
Abstract
Eight synthetic food colorants (Amaranth, Brilliant Blue, Indigo Carmine, New Red, Ponceau 4R, Sunset Yellow, Tartrazine, Allura Red) were determined by high-performance ion chromatography on an anion-exchange analytical column with very low hydrophobicity and visible absorbance detection. Gradient elution with hydrochloric acid-acetonitrile effected both the chromatographic separation of these colorants and the on-line clean-up of the analytical column, which was very advantageous for routine analysis. High-performance ion chromatography may be a solution to the chromatographic analysis for some water-soluble, organic analytes with strong hydrophobicity. The method has been applied to the determination of colorants in drinks and in instant drink powder. No time-consuming pretreatment, as used in conventional liquid chromatography, was needed.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9894346 DOI: 10.1016/s0021-9673(98)00759-6
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759