[Effects of a 6F urethral catheter on uroflowmetry in men with benign prostatic hypertrophy].

J F Hermieu1, V Ravery, R Le Coent, V Delmas, L Boccon-Gibod.   

Abstract

OBJECTIVE: To determine whether the presence of a 6F catheter in the urethra modifies uroflowmetric parameters in a population of men with benign prostatic hyperplasia. PATIENTS AND METHODS: 260 patients, 45 to 86 years old, presenting with benign prostatic hyperplasia, underwent two uroflowmetries, one with the urethra free, the other with a 6F urethral catheter. Patients with a history of neurological disease, diabetes mellitus, prostatic cancer, lower urinary tract surgery or taking drugs likely to affect micturition, were excluded from the study. Maximum flow rate (Qmax) and voided volume were then compared during the 2 recordings, using Wilcoxon's non parametric test.
RESULTS: No significant difference was observed between voided volumes in the 2 groups (m: 207 ml versus m: 226 ml), but a very significant difference was observed for maximum flow rate measurements (m: 8.62 ml/s versus m: 6.97 ml/s p = 0.0001).
CONCLUSION: The presence of an urethral catheter, even small calibre (6F), modifies the flow rate by more than 1.5 ml/s. This element must be taken into account when performing and interpreting measurements of the pressure-flow relationship in men.

Entities:  

Mesh:

Year:  1998        PMID: 9894264

Source DB:  PubMed          Journal:  Prog Urol        ISSN: 1166-7087            Impact factor:   0.915


  1 in total

1.  The effect of a 6 Fr catheter on flow rate in men.

Authors:  Patrick Richard; Nydia Icaza Ordonez; Le Mai Tu
Journal:  Urol Ann       Date:  2013-10
  1 in total