| Literature DB >> 9892084 |
M Dachtler1, K Kohler, K Albert.
Abstract
An efficient reversed-phase high-performance liquid chromatographic assay with ultraviolet detection at 450 nm for simultaneous determination of lutein and zeaxanthin stereoisomers in bovine retina is described. The procedure involves rapid and careful one-step hexane extraction of the carotenoids from the homogenized liquid and enrichment by on-line solid phase extraction on a polystyrene cartridge. The substances were eluted at a flow rate of 1 ml/min with acetone-water (85:15, v/v) on a C30 reversed-phase column. C30 phases exhibit superior shape selectivity for the separation of carotenoid stereoisomers compared with conventional C18 phases. For comparison a mixture of iodine-isomerized lutein and zeaxanthin standards was separated.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9892084 DOI: 10.1016/s0378-4347(98)00431-9
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273