Midplane dose determination during total body irradiation using in vivo dosimetry.

M Ribas1, N Jornet, T Eudaldo, D Carabante, M A Duch, M Ginjaume, G Gómez de Segura, F Sánchez-Doblado.   

Abstract

BACKGROUND AND
PURPOSE: During TBI techniques an accurate determination of the dose distribution is very difficult when using commercial treatment planning systems. In order to determine the midplane dose, an algorithm was developed based on the use of in vivo dosimetry.
MATERIALS AND METHODS: Scanditronix EDP-30 diodes were placed at the entrance and the exit surface for in vivo dosimetry. The proposed algorithm was validated firstly in a regular and homogeneous phantom of different thickness with an ionization chamber and TL dosimeters and secondly in an Alderson anthropomorphic phantom with TL dosimeters. In this study, in vivo measurements were evaluated in 60 patients and furthermore, in 20 of them, the midplane dose calculated with this algorithm was compared with the method described by Rizzotti A, Compri C, Garusi GF. Dose evaluation to patients irradiated by 60Co beams, by means of direct measurement on the incident and on the exit surfaces. Radiother. Oncol. 1985;3:279-283.
RESULTS: No differences were found between the two methods. The differences between dose values calculated with both methods and dose values measured with the ionization chamber and TL dosimeters were within +/-22% and +/-4%, respectively, in the regular and homogeneous phantom and within +/-2% in the Alderson phantom. The algorithm was useful in calculating the midplane dose when heterogeneities as lungs were present. Even when partial transmission blocks were used to reduce the dose to the lungs, the algorithm with modified correction factors gave a midplane lung dose in the Alderson phantom within 1.3% of the measurements with TL dosimeters. For 360 patients' measurements in each A-P and P-A field, the relative deviations were analyzed between the measured and calculated entrance, exit dose and midplane dose and the prescribed dose, always applying the temperature correction factor. These deviations at the entrance dose were within +/-4%. Greater deviations were found for the exit dose measurements. Deviations larger than +/-10% corresponded in general to obese patients, with a thickness over 25 cm. The relative deviations between the total received and prescribed midplane doses in 60 patients were within +/-3%.
CONCLUSIONS: The results indicate excellent correspondence between the total prescribed and calculated midplane doses using this algorithm while also no significant differences were found when the Rizzotti method was used. Comparison between doses measured with TL dosimeters in the core of Alderson phantom lungs and doses calculated from in vivo measurements showed that the proposed algorithm could be used in the presence of heterogeneities even when partial transmission blocks were used. The temperature correction factor must be applied in order to avoid a 2-3% dose overestimation.

Entities:  

Mesh:

Year:  1998        PMID: 9886703     DOI: 10.1016/s0167-8140(98)00094-2

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  4 in total

1.  Dosimetry and verification of Co total body irradiation with human phantom and semiconductor diodes.

Authors:  Mahmoud Allahverdi; Ghazale Geraily; Mahbod Esfehani; Aliakbar Sharafi; Peyman Haddad; Alireza Shirazi
Journal:  J Med Phys       Date:  2007-10

2.  Patient dose analysis in total body irradiation through in vivo dosimetry.

Authors:  K Ganapathy; P G G Kurup; V Murali; M Muthukumaran; N Bhuvaneshwari; J Velmurugan
Journal:  J Med Phys       Date:  2012-10

Review 3.  In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Igor Olaciregui-Ruiz; Sam Beddar; Peter Greer; Nuria Jornet; Boyd McCurdy; Gabriel Paiva-Fonseca; Ben Mijnheer; Frank Verhaegen
Journal:  Phys Imaging Radiat Oncol       Date:  2020-08-29

4.  In vivo dosimetry for total body irradiation: five-year results and technique comparison.

Authors:  Reshma P Patel; Alison J Warry; David J Eaton; Christopher H Collis; Ivan Rosenberg
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.