Literature DB >> 9886605

Rapid detection of elevated prostate-specific antigen levels in blood: performance of various membrane strip tests compared.

K Jung1, J Zachow, M Lein, B Brux, P Sinha, S Lenk, D Schnorr, S A Loening.   

Abstract

OBJECTIVES: To compare the ability of four commercially available membrane strip tests to detect increased (4 microg/L or more) concentrations of prostate-specific antigen (PSA) in blood.
METHODS: Serum samples with PSA concentrations less than 4 microg/L (n = 67) and from greater than 4 microg/L to 20 microg/L (n = 32) were independently examined by two observers using the PSA membrane strip tests from Chembio, Medpro, Seratec, and Syntron. The positive and negative results of each membrane strip test were classified as either true positive or negative and false negative or positive by comparing them with the quantitative PSA assay of Immulite DPC using the conventional threshold value of 4 microg/L.
RESULTS: The interobserver variations of the tests were between 93% and 97%. The color stability of the Seratec and Chembio tests did not show significant differences between test results read within 10 to 20 minutes of the reaction time; however, the results of the other two tests were especially affected by variations in the reading time. The sensitivity and specificity of the tests in relation to the threshold of 4 microg/L were 67% to 93% and 87% to 97%, respectively.
CONCLUSIONS: The Syntron test and, within certain limitations, the Seratec test fulfill the concept of a rapid and convenient PSA determination to detect PSA concentrations greater than 4 microg/L. Methodologic optimization of the tests by a grading of the PSA measuring ranges (eg, between 0 and 2, 3 and 4, 4 and 6, and 7 and 10 microg/L) should be taken into account for future development.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 9886605     DOI: 10.1016/s0090-4295(98)00419-1

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  4 in total

1.  Claros System: A Rapid Microfluidics-Based Point-of-Care System for Quantitative Prostate Specific Antigen Analysis from Finger-Stick Blood.

Authors:  Agnieszka Maj-Hes; Sabina Sevcenco; Tibor Szarvas; Gero Kramer
Journal:  Adv Ther       Date:  2019-02-18       Impact factor: 3.845

2.  Quantitation of femtomolar protein levels via direct readout with the electrochemical proximity assay.

Authors:  Jiaming Hu; Tanyu Wang; Joonyul Kim; Curtis Shannon; Christopher J Easley
Journal:  J Am Chem Soc       Date:  2012-04-10       Impact factor: 15.419

3.  Evaluation of a rapid quantitative determination method of PSA concentration with gold immunochromatographic strips.

Authors:  Cheng-Ching Wu; Hung-Yu Lin; Chao-Ping Wang; Li-Fen Lu; Teng-Hung Yu; Wei-Chin Hung; Jer-Yiing Houng; Fu-Mei Chung; Yau-Jiunn Lee; Jin-Jia Hu
Journal:  BMC Urol       Date:  2015-11-03       Impact factor: 2.264

4.  Evaluation of a rapid one-step PSA test for primary prostate cancer screening.

Authors:  Shingo Ashida; Ichiro Yamasaki; Chiaki Kawada; Hideo Fukuhara; Satoshi Fukata; Kenji Tamura; Takashi Karashima; Keiji Inoue; Taro Shuin
Journal:  BMC Urol       Date:  2021-09-27       Impact factor: 2.264