Four-year experience with the routine use of the programmable Hakim valve in the management of children with hydrocephalus.

V Rohde1, L Mayfrank, V T Ramakers, J M Gilsbach.   

Abstract

OBJECTIVE: Cerebrospinal fluid (CSF) over- and underdrainage symptoms are frequent sequelae of shunt placement in patients with hydrocephalus, sometimes requiring repeated operations. To achieve more adequate CSF drainage, the non-invasively programmable Hakim valve has been developed. Because the clinical experiences with this valve so far are confined to adults, we describe our experiences with the routine use of the programmable Hakim valve in childhood hydrocephalus.
METHOD: Sixty children (mean age of 3.4 years) with hydrocephalus of various aetiologies have been shunted with the programmable Hakim valve. In the majority of cases, initial opening pressures of between 100 and 120 mm H2O were selected. The mean follow-up period was 2.1 years.
RESULTS: Thirty-three readjustment of the pressure setting of the valve were performed in 20 children because of CSF overdrainage (low intracranial pressure syndrome n = 13, slit ventricle syndrome n = 2, hygroma n = 1), CSF underdrainage (n = 3) and CSF leakage through the operation wound (n = 1). The symptoms of inadequate CSF drainage were cured in 18 of the 20 children. The necessity for valve readjustments was independent of the aetiology of the hydrocephalus. Thirty-one complications requiring repeated operation occurred during the follow-up period, accounting for an annual complication rate of 24.6%. Three complications were valve-related.
CONCLUSION: In the majority of cases, the programmable Hakim valve allows the successful management of symptoms related to CSF over- and underdrainage by non-invasive change of the initial pressure setting of the valve. Therefore, the programmable Hakim valve should be considered as an alternative to non-programmable valves of advanced design.

Entities:  

Mesh:

Year:  1998        PMID: 9870057     DOI: 10.1007/s007010050226

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  11 in total

1.  [Gravity valves for idiopathic normal pressure hydrocephalus. A Prospective study of 60 patients].

Authors:  U Meier
Journal:  Nervenarzt       Date:  2004-06       Impact factor: 1.214

2.  The Strata programmable valve for shunt-dependent hydrocephalus: the pediatric experience at a single institution.

Authors:  Edward S Ahn; Markus Bookland; Benjamin S Carson; Jon D Weingart; George I Jallo
Journal:  Childs Nerv Syst       Date:  2006-10-07       Impact factor: 1.475

Review 3.  Shunt overdrainage syndrome: review of the literature.

Authors:  Bienvenido Ros; Sara Iglesias; Álvaro Martín; Antonio Carrasco; Guillermo Ibáñez; Miguel A Arráez
Journal:  Neurosurg Rev       Date:  2017-03-29       Impact factor: 3.042

4.  Using individualized opening pressure to determine the optimal setting of an adjustable proGAV shunt in treatment of hydrocephalus in infants.

Authors:  Li Xinxing; Diao Hongyu; Liu Yunhui
Journal:  Childs Nerv Syst       Date:  2015-07-05       Impact factor: 1.475

Review 5.  Is there an ideal shunt? A panoramic view of 110 years in CSF diversions and shunt systems used for the treatment of hydrocephalus: from historical events to current trends.

Authors:  Nigel Peter Symss; Shizuo Oi
Journal:  Childs Nerv Syst       Date:  2014-12-30       Impact factor: 1.475

6.  Adjustments in gravitational valves for the treatment of childhood hydrocephalus-a retrospective survey.

Authors:  Anna Felicitas Gebert; Matthias Schulz; Hannes Haberl; Ulrich-Wilhelm Thomale
Journal:  Childs Nerv Syst       Date:  2013-05-29       Impact factor: 1.475

7.  Management of neonatal hydrocephalus: feasibility of use and safety of two programmable (Sophy and Polaris) valves.

Authors:  Juan F Martínez-Lage; María-José Almagro; Isabel Sanchez Del Rincón; Miguel A Pérez-Espejo; Claudio Piqueras; Raúl Alfaro; Javier Ros de San Pedro
Journal:  Childs Nerv Syst       Date:  2007-10-09       Impact factor: 1.475

8.  A pressure adjustment protocol for programmable valves.

Authors:  Kyoung-Hun Kim; In-Seoung Yeo; Jin-Seok Yi; Hyung-Jin Lee; Ji-Ho Yang; Il-Woo Lee
Journal:  J Korean Neurosurg Soc       Date:  2009-10-31

9.  Cerebrospinal Fluid Lumbar Tapping Utilization for Suspected Ventriculoperitoneal Shunt Under-Drainage Malfunctions.

Authors:  Jong-Beom Lee; Ho-Young Ahn; Hong-Jae Lee; Ji-Ho Yang; Jin-Seok Yi; Il-Woo Lee
Journal:  J Korean Neurosurg Soc       Date:  2016-12-29

10.  A review of the current treatment methods for posthaemorrhagic hydrocephalus of infants.

Authors:  David Shooman; Howard Portess; Owen Sparrow
Journal:  Cerebrospinal Fluid Res       Date:  2009-01-30