Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings.

D V Exner1, J M Rothschild, S Heal, A M Gillis.   

Abstract

Bipolar lead use has increased due to oversensing concerns with older unipolar systems. Data on contemporary unipolar devices with improved hardware design and greater programming flexibility is lacking. Using a randomized crossover design, unipolar and bipolar sensing characteristics of 22 atrial and 16 ventricular leads were compared in 34 patients who had pulse generators of programmable polarity. Unipolar and bipolar intracardiac electrogram amplitudes, pacing and sensing thresholds at rest were similar. Provocative maneuvers were used to assess for myopotential inhibition. At atrial sensitivities of 0.625-1.50 mV, myopotential inhibition occurred in 11 (50%) atrial leads in the unipolar mode compared to 1 (5%) in the bipolar mode (p < 0.001). At sensitivities of > 1.50 mV myopotential inhibition occurred in only 1 ventricular (unipolar) lead. An optimal sensitivity setting for each polarity was derived using clinic test results and assessed by ambulatory ECG (AECG). At these optimal settings, oversensing occurred in 1 (6%) atrial and 1 (8%) ventricular unipolar lead during AECG monitoring, whereas oversensing was not seen in any leads programmed to the bipolar mode. Undersensing occurred in 5 (29%) atrial unipolar versus 1 (6%) bipolar lead (p = 0.08). Undersensing was not observed in any of the ventricular leads. Myopotential inhibition may be frequently provoked by provocative maneuvers at higher sensitivity settings in atrial unipolar leads. The frequency of oversensing can be significantly reduced by defining an optimal sensitivity setting using simple isometric maneuvers. Given present day concerns over bipolar lead longevity, increased utilization of unipolar ventricular leads should be considered.

Entities:  

Mesh:

Year:  1998        PMID: 9869994     DOI: 10.1023/a:1009708606035

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  23 in total

1.  Characteristics and clinical effects of myopotential signals in a unipolar DDD pacemaker population.

Authors:  S H Zimmern; M F Clark; W K Austin; J M Fedor; J J Gallagher; R H Svenson; J L Duncan
Journal:  Pacing Clin Electrophysiol       Date:  1986-11       Impact factor: 1.976

2.  Reduced incidence of myopotential pacemaker inhibition by abdominal generator implantation.

Authors:  M Rosenqvist; R Nordlander; M Andersson; O Edhag
Journal:  Pacing Clin Electrophysiol       Date:  1986-05       Impact factor: 1.976

3.  Survey of cardiac pacing and defibrillation in the United States in 1993.

Authors:  A D Bernstein; V Parsonnet
Journal:  Am J Cardiol       Date:  1996-07-15       Impact factor: 2.778

4.  Comparison of myopotential interference in unipolar-bipolar programmable DDD pacemakers.

Authors:  M D Gabry; M Behrens; C Andrews; M Wanliss; P T Klementowicz; S Furman
Journal:  Pacing Clin Electrophysiol       Date:  1987-11       Impact factor: 1.976

5.  Interference effect of myopotenitals on function of unipolar demand pacemakers.

Authors:  O J Ohm; H Bruland; O M Pedersen; E Waerness
Journal:  Br Heart J       Date:  1974-01

6.  Inhibition of demand pacemakers by skeletal muscle potentials.

Authors:  D A Winter
Journal:  JAMA       Date:  1973-01-29       Impact factor: 56.272

7.  Unipolar sensing abnormalities: incidence and clinical significance of skeletal muscle interference and undersensing in 228 patients.

Authors:  S I Secemsky; R G Hauser; P Denes; L M Edwards
Journal:  Pacing Clin Electrophysiol       Date:  1982-01       Impact factor: 1.976

Review 8.  Pacemaker therapy in patients with atrial fibrillation.

Authors:  A Pollak; R H Falk
Journal:  Am Heart J       Date:  1993-03       Impact factor: 4.749

9.  The clinical relevance of electromyopotential oversensing in current unipolar devices.

Authors:  J N Gross; S Platt; R Ritacco; C Andrews; S Furman
Journal:  Pacing Clin Electrophysiol       Date:  1992-11       Impact factor: 1.976

10.  Multicenter experience with a bipolar tined polyurethane ventricular lead.

Authors:  D L Hayes; K J Graham; M Irwin; H Vidaillet; G Disler; M Sweesy; D Kincaid; M J Osborn; V J Suman; S A Neubauer
Journal:  Pacing Clin Electrophysiol       Date:  1995-05       Impact factor: 1.976

View more
  1 in total

1.  [Sport for pacemaker patients].

Authors:  C W Israel
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.