| Literature DB >> 9869380 |
J R Veraart1, C Gooijer, H Lingeman, N H Velthorst, U A Brinkman.
Abstract
The analysis of complex biological samples with capillary electrophoresis (CE) requires proper sample pretreatment. In this paper the applicability of solid-phase extraction (SPE) coupled at-line with CE is studied, by using a laboratory-made interface. A fresh (disposable) SPE cartridge is used for each sample to prevent carry-over effects. The sample handling procedure is performed parallel with the analysis of the previous sample, to improve sample throughput. Using this set-up, negatively charged test compounds (some non-steroid anti-inflammatory drugs) can be determined in serum and urine. The method is linear over at least two decades and detection limits are around 40 microg/l. A single capillary, flushed only once a week with a sodium hydroxide solution, was used without problems for the analysis of ca. 900 samples during 1 year. The robustness of the system was very good: no blocking of loop, interface or capillary was found during this period. Furthermore, the system was successfully used for overnight runs.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9869380 DOI: 10.1016/s0378-4347(98)00410-1
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273