Abstract

The variations in dose output with collimator jaw setting have been investigated for treatment fields shaped by a multileaf collimator. Measurements have been made for 6 and 15 MV x-ray beams on a Varian Clinac 2100C machine. The results of our study show that the collimator jaw settings can affect the dose by about 5% for a small field shaped by a multileaf collimator. The effect is smaller for larger fields.

Mesh:

Year:  1998        PMID: 9863730     DOI: 10.1016/s0958-3947(98)00031-4

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482