| Literature DB >> 9863730 |
Abstract
The variations in dose output with collimator jaw setting have been investigated for treatment fields shaped by a multileaf collimator. Measurements have been made for 6 and 15 MV x-ray beams on a Varian Clinac 2100C machine. The results of our study show that the collimator jaw settings can affect the dose by about 5% for a small field shaped by a multileaf collimator. The effect is smaller for larger fields.Mesh:
Year: 1998 PMID: 9863730 DOI: 10.1016/s0958-3947(98)00031-4
Source DB: PubMed Journal: Med Dosim ISSN: 1873-4022 Impact factor: 1.482