Literature DB >> 9856706

Circulating blood volume measured by pulse dye-densitometry: comparison with (131)I-HSA analysis.

T Iijima1, Y Iwao, H Sankawa.   

Abstract

BACKGROUND: Pulse dye-densitometry (PDD) is a newly developed technique for monitoring the arterial concentration of indocyanine green. Using this method, circulating blood volume (CBV) can be calculated without using radioisotopes. In this study, the CBV value obtained by PDD was validated by comparison using the human serum albumin ((131)I-HSA) dilution method.
METHODS: Eleven healthy volunteers underwent placement of cannulae into the radial artery and antecubital vein for withdrawal of blood samples and injection of indicator. Probes for PDD were attached to the right nostril and the right index finger. Indocyanine green (20 mg), dissolved in 4 ml water, and 25 microCi (131)I-HSA in 1 ml distilled water were injected simultaneously into the left antecubital vein. Blood samples were withdrawn 3, 6, 10, 20, 30, and 45 min after injection, then processed for spectrophotometric measurement of indocyanine green and scintillation counting.
RESULTS: The blood dye concentration correlated well with the values obtained by PDD (r=0.986, imprecision 0.04+/-0.11 mg/l, 10.0+/-31%. The imprecision of the CBV value obtained by PDD (nose probe) and by the (131)I-HSA dilution method was 3.99+/-10.54%, 0.259+/-0.593 l. The imprecision of the CBV obtained by in vitro spectrophotometry compared with PDD was 2.47+/-9.00%, 0.100+/-0.446 l.
CONCLUSIONS: This newly developed, less invasive method can measure CBV with an imprecision of 3.99+/-10.54%, 0.259+/-0.593 l (nose probe), and thus is also as accurate as the conventional radioisotope method.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9856706     DOI: 10.1097/00000542-199812000-00009

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  17 in total

1.  Monitoring techniques for blood volume in the anethesia and intensive care setting.

Authors:  B Allaria; M Favaro; M Resta
Journal:  J Clin Monit Comput       Date:  2002 Apr-May       Impact factor: 2.502

2.  Closed-Loop Intravenous Drug Administration Using Photoplethysmography.

Authors:  George W Carpenter; Holly G Myers; Eric A Sherer; Katie A Evans; D Patrick O'Neal
Journal:  IEEE J Transl Eng Health Med       Date:  2018-11-01       Impact factor: 3.316

3.  Determination of the standard value of circulating blood volume during anesthesia using pulse dye-densitometry: a multicenter study in Japan.

Authors:  Takehiko Iijima; Hiroshi Ueyama; Yoshiyuki Oi; Isao Fukuda; Hironori Ishihara; Hikaru Kohase; Yoshifumi Kotake; Kaoru Koyama; Hideki Miyao; Naoki Kobayashi
Journal:  J Anesth       Date:  2005       Impact factor: 2.078

Review 4.  [Determinants of insensible fluid loss. Perspiration, protein shift and endothelial glycocalyx].

Authors:  M Jacob; D Chappell; K Hofmann-Kiefer; P Conzen; K Peter; M Rehm
Journal:  Anaesthesist       Date:  2007-08       Impact factor: 1.041

Review 5.  Complexity of blood volume control system and its implications in perioperative fluid management.

Authors:  Takehiko Iijima
Journal:  J Anesth       Date:  2009-11-18       Impact factor: 2.078

6.  Comparison of markers of circulating blood volume in hemodialysis patients.

Authors:  Yoshiyuki Morishita; Yasuhiro Ando; Eriko Ishii; Mayumi Arisaka; Eiji Kusano
Journal:  Clin Exp Nephrol       Date:  2005-09       Impact factor: 2.801

7.  Quantitative measurement of lymphatic function in mice by noninvasive near-infrared imaging of a peripheral vein.

Authors:  Steven T Proulx; Qiaoli Ma; Diana Andina; Jean-Christophe Leroux; Michael Detmar
Journal:  JCI Insight       Date:  2017-01-12

8.  Stability of cirrhotic systemic hemodynamics ensures sufficient splanchnic blood flow after living-donor liver transplantation in adult recipients with liver cirrhosis.

Authors:  Tomohide Hori; Shintaro Yagi; Taku Iida; Kentaro Taniguchi; Kentaro Yamagiwa; Chiduru Yamamoto; Takashi Hasegawa; Koichiro Yamakado; Takuma Kato; Kanako Saito; Linan Wang; Mie Torii; Yukinobu Hori; Kan Takeda; Kazuo Maruyama; Shinji Uemoto
Journal:  World J Gastroenterol       Date:  2007-11-28       Impact factor: 5.742

Review 9.  Systemic hemodynamics in advanced cirrhosis: Concerns during perioperative period of liver transplantation.

Authors:  Tomohide Hori; Yasuhiro Ogura; Yasuharu Onishi; Hideya Kamei; Nobuhiko Kurata; Motoshi Kainuma; Hideo Takahashi; Shogo Suzuki; Takashi Ichikawa; Shoko Mizuno; Tadashi Aoyama; Yuki Ishida; Takahiro Hirai; Tomoko Hayashi; Kazuko Hasegawa; Hiromu Takeichi; Atsunobu Ota; Yasuhiro Kodera; Hiroyuki Sugimoto; Taku Iida; Shintaro Yagi; Kentaro Taniguchi; Shinji Uemoto
Journal:  World J Hepatol       Date:  2016-09-08

10.  Blood volume measurement with indocyanine green pulse spectrophotometry: dose and site of dye administration.

Authors:  Menno R Germans; Philip C de Witt Hamer; Leonard J van Boven; Koos A H Zwinderman; Gerrit J Bouma
Journal:  Acta Neurochir (Wien)       Date:  2009-09-04       Impact factor: 2.216