| Literature DB >> 9855223 |
V Saikko1, J Nevalainen, H Revitzer, P Ylinen.
Abstract
We used a hip joint simulator to compare the metal release from CoCr/CoCr, CoCr/PE, and alumina/PE total hip articulations. The metal release was quantified by analyzing the Co, Cr, and Ni contents of the bovine serum lubricant used with atomic absorption spectroscopy. CoCr/CoCr articulations released substantial amounts of metal, whereas CoCr/PE was equal to the control, alumina/PE, in that metal release was negligible. The metal release was in accordance with the known clinical wear rates of CoCr/CoCr articulations. The largest dimensional changes occurred in polyethylene cups, the penetrations of CoCr heads to the polyethylene cups being twice that of the alumina head, which is consistent with clinical experience. The research on the wear behavior of different materials, aiming to find a prosthesis with negligible wear, needs to be continued. Due to the substantial metal release, the CoCr/CoCr articulation is hardly the final solution of the wear problem in total hip arthroplasty.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9855223 DOI: 10.3109/17453679808997777
Source DB: PubMed Journal: Acta Orthop Scand ISSN: 0001-6470