R T Andrews1, J F Geschwind, S J Savader, A C Venbrux.
Abstract
We present four patients in whom bedside placement of a central venous catheter was complicated by entrapment of a J-tip guidewire by a previously placed vena cava (VC) filter. Two Venatech filters were fragmented and displaced into the superior VC or brachiocephalic vein during attempted withdrawal of the entrapped wire. Two stainless-steel Greenfield filters remained in place and intact. Fluoroscopically guided extraction of both wires entrapped by Greenfield filters was successfully performed in the angiography suite.Entities: