Tumour dose estimation using automated TLD techniques.

H M Ferguson1, G D Lambert, D Gustard, R M Harrison.   

Abstract

Lithium fluoride (TLD-700) dosimeters were used to measure exit surface absorbed doses in external beam radiotherapy using an automated TLD reader. Delivered tumour absorbed doses were derived from these measurements for head and neck, pelvis and breast treatments. For the head and neck treatments (first fraction only), the mean percentage difference between prescribed and delivered tumour absorbed doses was -0.15 +/- 3.0% (+/- 1 SD), for the pelvic treatments -0.83 +/- 2.8% and for the breast treatments +0.26 +/- 2.9%. The spread of results is approximately +/- 3% (+/- 1 SD). This is comparable with the estimated uncertainty in a single TLD absorbed dose measurement in phantom (+/- 2%; +/- 1 SD). Thus, ICRU recommended tolerances for absorbed dose delivery of +/- 5% may not be unequivocally detectable using this method. An action level of +/- 10% is suggested, allowing investigation of possible gross errors in treatment delivery at an early stage, before the course of treatment has progressed to a point at which absorbed dose compensation is impossible.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9831378     DOI: 10.1080/028418698430449

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  2 in total

1.  Verification of Entrance Dose Measurements with Thermoluminescent Dosimeters in Conventional Radiotherapy Procedures Delivered with Co-60 Teletherapy Machine.

Authors:  O B Evwierhurhoma; Z A Ibitoye; C A Ojieh; Jtk Duncan
Journal:  Ann Med Health Sci Res       Date:  2015 Nov-Dec

2.  Evaluation of target dose based on water-equivalent thickness in external beam radiotherapy.

Authors:  Behnaz Ghanbar Moghaddam; Masoud Vahabi-Moghaddam; Alireza Sadremomtaz
Journal:  J Med Phys       Date:  2013-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.