| Literature DB >> 9823119 |
A Busslinger1, B Sener, F Barbakow.
Abstract
Sodium hypochlorite is a major irrigant in endodontics, and nickel-titanium instruments are gaining in popularity. This paper investigated the corrosion of nickel-titanium Lightspeed instruments in 1% and 5% NaOCl solutions. The instruments were immersed in ultrasonicated NaOCl solutions for varying times up to 1 h. Corrosion was determined by electrothermal absorption spectrometry in 100 microL aliquots of NaOCl. Background contamination of nickel in the 1% and 5% NaOCl solutions used was low, but high enough to interfere in detecting any increases in nickel after immersing the instruments. The amounts of titanium recorded in the 1% NaOCl solutions were insignificant. However, a statistically significant amount of titanium was detected from the Lightspeed instruments after immersion times of 30 and 60 min in 5% NaOCl. Clinically such instruments do not have an 'in situ' time of 30 min, and this corrosion may be considered irrelevant clinically.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9823119 DOI: 10.1046/j.1365-2591.1998.00149.x