Literature DB >> 9823115

Shaping ability of Quantec Series 2000 rotary nickel-titanium instruments in simulated root canals: Part 1.

S A Thompson1, P M Dummer.   

Abstract

The aim of this study was to determine the shaping ability of Quantec Series 2000 nickel-titanium instruments in simulated canals. A total of 40 simulated root canals made up of four different shapes in terms of angle and position of curvature were prepared by Quantec instruments using the technique recommended by the manufacturer. Part 1 of this two-part report describes the efficacy of the instruments in terms of preparation time, instrument failure, canal blockages, change in canal length and three-dimensional canal form. The time necessary for canal preparation was on average 5.7 min and was significantly influenced (P < 0.01) by canal shape. One instrument fractured and three size nine instruments deformed; however, canal shape did not influence significantly instrument failure. All of the canals remained patent, none became blocked with debris. The majority of the canals maintained working distance (26 out of 40), however the mean change in length differed significantly (P < 0.05) between canal types. Overall, canals with 40 degrees curves lost length whilst those with 20 degrees curves gained in length. Examination of intracanal impressions revealed that preparation with Quantec Series 2000 instruments produced canals with definite apical stops, smooth canal walls and good flow and taper. However, the quality of apical smoothness and flow was influenced significantly (P < 0.0001) by canal shape with specimens having 40 degrees canals displaying less desirable qualities. Under the conditions of this study, Quantec Series 2000 rotary nickel-titanium instruments prepared simulated canals rapidly, safely and with good three-dimensional form.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9823115     DOI: 10.1046/j.1365-2591.1998.00151.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  3 in total

1.  Comparison of Canal Preparation Pattern of K3 and ProTaper Rotary Files in Curved Resin Blocks.

Authors:  Nahid Mohammadzade Akhlaghi; Zohreh Khalilak; Ladan Baradaran Mohajeri; Mahshid Sheikholeslami; Saeed Saedi
Journal:  Iran Endod J       Date:  2008-04-02

Review 2.  Rotary endodontics in primary teeth - A review.

Authors:  Sageena George; S Anandaraj; Jyoti S Issac; Sheen A John; Anoop Harris
Journal:  Saudi Dent J       Date:  2015-11-22

3.  A comparative study on the shaping ability of k3, profile and protaper instruments in simulated curved root canals.

Authors:  Chinni Suneelkumar; Chellaswamy Savarimalai Karumaran; Sundararaman Ramachandran; Rajamani Indira; Padmabhan Shankar; Anil Kumar
Journal:  Iran Endod J       Date:  2010-08-15
  3 in total