Literature DB >> 9818343

Evaluation of a 3-piece silicone intraocular lens with poly(methyl methacrylate) haptics.

R J Linnola1, A Holst.   

Abstract

PURPOSE: To evaluate the safety and performance of a foldable silicone intraocular lens (IOL) in small incision phacoemulsification surgery.
SETTING: Vaasa Central Hospital, Vaasa, Finland.
METHODS: Phacoemulsification with primary implantation of a posterior chamber IOL was performed in 50 patients. All eyes were implanted with a CeeOn IOL (model 920, Pharmacia & Upjohn) with a silicone optic and poly(methyl methacrylate) haptics. The ease of folding, implanting, and unfolding the IOL in the bag were recorded. Visual acuity, postoperative complications, optic decentration, and optic tilt were measured at 5 postoperative visits. The IOLs were evaluated for haze and discoloration and the IOL surface, for inflammatory or pigment deposits. Follow-up was 12 months.
RESULTS: In 44 of 50 cases, folding the IOL was easy, and in 46 of 50, unfolding the IOL was controlled. The mean follow-up was 11.45 months +/- 0.52 (SD). No postoperative complications occurred, and 96% of patients achieved a best corrected visual acuity (BCVA) of 0.5 or better and 56%, a BCVA of 1.0 or better. The IOL decentration was less than 0.25 mm in 89% of cases, and there was no optic tilt. The neodymium:YAG laser capsulotomy rate was 11%.
CONCLUSION: Good visual outcome and excellent centration were achieved with the CeeOn model 920 IOL. It is therefore suitable for small incision cataract surgery. Because of the IOL's smooth, polished surface, the choice of instruments for folding and implantation is important.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9818343     DOI: 10.1016/s0886-3350(98)80175-2

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Comparison of outcomes with multifocal intraocular lenses: a meta-analysis.

Authors:  Béatrice Cochener; Antoine Lafuma; Babak Khoshnood; Laurène Courouve; Gilles Berdeaux
Journal:  Clin Ophthalmol       Date:  2011-01-07

Review 2.  Research Progress Concerning a Novel Intraocular Lens for the Prevention of Posterior Capsular Opacification.

Authors:  Yidong Zhang; Chengshou Zhang; Silong Chen; Jianghua Hu; Lifang Shen; Yibo Yu
Journal:  Pharmaceutics       Date:  2022-06-25       Impact factor: 6.525

3.  Modelling lifetime cost consequences of ReSTOR in cataract surgery in four European countries.

Authors:  Antoine Lafuma; Gilles Berdeaux
Journal:  BMC Ophthalmol       Date:  2008-07-15       Impact factor: 2.209

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.