A D Hanlon1, S J Gibbs, L D Hall, D E Haycock, W J Frith, S Ablett.
Abstract
A narrow-gap, temperature-controlled Couette flow rheometer has been developed to study fluid velocities within the annular gap between two concentric cylinders by nuclear magnetic resonance (NMR) imaging and velocimetry. Alternative pulsed-field-gradient-based nuclear magnetic resonance imaging strategies which may be used for measurement of velocity within the Couette flow device have been evaluated. These include two-dimensional (2-D) imaging techniques with acquisition times of several minutes and a one-dimensional (1-D) projection method which exploits the symmetry of the device to reduce overall measurement time to less than 1 min. Velocity measurements made using each technique are presented for a Newtonian fluid undergoing Couette flow at shear rates of approximately 20 and 60 s(-1).Mesh:
Year: 1998 PMID: 9814778 DOI: 10.1016/s0730-725x(98)00089-7
Source DB: PubMed Journal: Magn Reson Imaging ISSN: 0730-725X Impact factor: 2.546