Improving calibration accuracy in gel dosimetry.

M Oldham1, M McJury, I B Baustert, S Webb, M O Leach.   

Abstract

A new method of calibrating gel dosimeters (applicable to both Fricke and polyacrylamide gels) is presented which has intrinsically higher accuracy than current methods, and requires less gel. Two test-tubes of gel (inner diameter 2.5 cm, length 20 cm) are irradiated separately with a 10 x 10 cm2 field end-on in a water bath, such that the characteristic depth-dose curve is recorded in the gel. The calibration is then determined by fitting the depth-dose measured in water, against the measured change in relaxivity with depth in the gel. Increased accuracy is achieved in this simple depth-dose geometry by averaging the relaxivity at each depth. A large number of calibration data points, each with relatively high accuracy, are obtained. Calibration data over the full range of dose (1.6-10 Gy) is obtained by irradiating one test-tube to 10 Gy at dose maximum (Dmax), and the other to 4.5 Gy at Dmax. The new calibration method is compared with a 'standard method' where five identical test-tubes of gel were irradiated to different known doses between 2 and 10 Gy. The percentage uncertainties in the slope and intercept of the calibration fit are found to be lower with the new method by a factor of about 4 and 10 respectively, when compared with the standard method and with published values. The gel was found to respond linearly within the error bars up to doses of 7 Gy, with a slope of 0.233 +/- 0.001 s(-1) Gy(-1) and an intercept of 1.106 +/- 0.005 Gy. For higher doses, nonlinear behaviour was observed.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9814511     DOI: 10.1088/0031-9155/43/10/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Characterization of a new radiochromic three-dimensional dosimeter.

Authors:  P Y Guo; J A Adamovics; M Oldham
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

2.  Echo-spacing optimization for the simultaneous measurement of reversible (R2') and irreversible (R2) transverse relaxation rates.

Authors:  Ruitian Song; Hee Kwon Song
Journal:  Magn Reson Imaging       Date:  2006-11-13       Impact factor: 2.546

3.  A practical three-dimensional dosimetry system for radiation therapy.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.