Literature DB >> 9806621

Reliability of phase-velocity measurements of tibial bone.

E D de Bruin1, R H Rozendal, E Stüssi.   

Abstract

BACKGROUND AND
PURPOSE: The Bone Stiffness Measurement Device-Swing is capable of measuring the propagation velocity of flexural waves in human tibial bone, which relates to bending stiffness. If the interrater and intrarater reliability of measurements obtained with the device are established, it can be used with confidence in assessing changes in bone. The purposes of this study were to detect potential sources of measurement error and to establish the interrater and intrarater reliability of measurements taken with the device. SUBJECTS AND METHODS: In the first part of the study, a random-effects design was used to obtain phase-velocity measurements in subjects without known orthopedic or neurological impairments. The second part of the study consisted of possible applications of the device with mixed designs on subjects with spinal cord injuries. By means of generalizability theory, multiple sources of error (eg, occasion, clinician, repetition) were estimated. For the clinical trial, 17 persons with spinal cord injuries not older than 5 weeks were tested.
RESULTS: The standard error of measurement (SEM) for intrarater reliability measurements ranged from 7.3 to 9.8 m x s(-1) . The SEMs for interrater reliability measurements ranged from 5.7 to 9.5 m x s(-1). The SEMs for measurements obtained by a single clinician in a clinical population ranged from 11.9 to 39.7 m x s(-1). CONCLUSION AND DISCUSSION: The reproducibility of measurements obtained with the device is suitably high for the device to be used for evaluation in clinical and research settings.

Entities:  

Mesh:

Year:  1998        PMID: 9806621     DOI: 10.1093/ptj/78.11.1166

Source DB:  PubMed          Journal:  Phys Ther        ISSN: 0031-9023


  5 in total

1.  Reliability of Phase Velocity Measurements of Flexural Acoustic Waves in the Human Tibia In-Vivo.

Authors:  Florian Vogl; Karin Schnüriger; Hans Gerber; William R Taylor
Journal:  PLoS One       Date:  2016-03-25       Impact factor: 3.240

2.  Hip Range of Motion and Strength in Male Athletes with Stage 1 Osteitis Pubis: A Cross-Sectional and Correlational Study.

Authors:  Luis Ceballos-Laita; Ignacio Hernando-Garijo; Ricardo Medrano-de-la-Fuente; María Teresa Mingo-Gómez; Andoni Carrasco-Uribarren; Sandra Jiménez-Del-Barrio
Journal:  Int J Environ Res Public Health       Date:  2022-10-06       Impact factor: 4.614

3.  Hand-held dynamometry in patients with haematological malignancies: measurement error in the clinical assessment of knee extension strength.

Authors:  Ruud H Knols; Geert Aufdemkampe; Eling D de Bruin; Daniel Uebelhart; Neil K Aaronson
Journal:  BMC Musculoskelet Disord       Date:  2009-03-09       Impact factor: 2.362

4.  Validity and reliability of accelerometer-based gait assessment in patients with diabetes on challenging surfaces.

Authors:  Eling D de Bruin; Michèle Hubli; Pamela Hofer; Peter Wolf; Kurt Murer; Wiebren Zijlstra
Journal:  J Aging Res       Date:  2012-07-31

5.  Discriminant validity and test re-test reproducibility of a gait assessment in patients with vestibular dysfunction.

Authors:  Annatina Schmidheiny; Jaap Swanenburg; Dominik Straumann; Eling D de Bruin; Ruud H Knols
Journal:  BMC Ear Nose Throat Disord       Date:  2015-10-22
  5 in total