New concept in echocardiography: harmonic imaging of tissue without use of contrast agent.

K Caidahl1, E Kazzam, J Lidberg, G Neumann Andersen, J Nordanstig, S Rantapää Dahlqvist, A Waldenström, R Wikh.   

Abstract

BACKGROUND: Endocardial border detection is important for echocardiographic assessment of left-ventricular function. Second harmonic imaging of contrast agents enhances this border detection. We discovered that harmonic imaging improves tissue visualisation even before contrast injection. We therefore sought objectively to demonstrate the degree of enhancement of endocardial and myocardial visualisation.
METHODS: An ATL HDI-3000 scanner with software for contrast harmonic imaging was used to record short-axis images of the left ventricle in 27 patients with possible myocardial disease and 22 controls, in the fundamental mode and with harmonic imaging. A computer program measured the relative grey-scale values within six segments of the endocardium and myocardium. An Acuson Sequoia scanner equipped with software for tissue harmonic imaging was used to investigate the reproducibility of ejection-fraction calculations in 22 patients with ischaemic heart disease.
FINDINGS: Harmonic imaging produced brighter endocardium within each segment. Relative to the mean grey value of the total imaging sector, the values for harmonic and fundamental imaging were 171.5 vs 85.6% (p<0.0001) in end diastole and 194.1 vs 106.7% (p<0.0001) in end systole. Results for the myocardial segments were also significantly better for harmonic imaging. Structure enhancement of similar magnitude was seen among patients and healthy controls. Use of harmonic imaging reduced the proportion of unacceptable images by 14-46% in different views and improved the reproducibility of biplane ejection-fraction measurements.
INTERPRETATION: In comparison with fundamental imaging, the relative endocardial and myocardial brightness is enhanced by harmonic imaging.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9788456     DOI: 10.1016/s0140-6736(98)02361-7

Source DB:  PubMed          Journal:  Lancet        ISSN: 0140-6736            Impact factor:   79.321


  12 in total

1.  Improved identification of viable myocardium using second harmonic imaging during dobutamine stress echocardiography.

Authors:  F B Sozzi; D Poldermans; J J Bax; A Elhendy; E C Vourvouri; R Valkema; J De Sutter; A F Schinkel; A Borghetti; J R Roelandt
Journal:  Heart       Date:  2001-12       Impact factor: 5.994

2.  Non-contrast second harmonic imaging improves interobserver agreement and accuracy of dobutamine stress echocardiography in patients with impaired image quality.

Authors:  A Franke; R Hoffmann; H P Kühl; W Lepper; O A Breithardt; M Schormann; P Hanrath
Journal:  Heart       Date:  2000-02       Impact factor: 5.994

Review 3.  Contrast Ultrasound, Sonothrombolysis and Sonoperfusion in Cardiovascular Disease: Shifting to Theragnostic Clinical Trials.

Authors:  Soufiane El Kadi; Thomas R Porter; Niels J W Verouden; Albert C van Rossum; Otto Kamp
Journal:  JACC Cardiovasc Imaging       Date:  2021-10-13

4.  Prevalence and development of additional cardiac abnormalities in 1448 patients with congenital ventricular septal defects.

Authors:  S Glen; J Burns; P Bloomfield
Journal:  Heart       Date:  2004-11       Impact factor: 5.994

5.  Long-term prognostic value of mitral regurgitation in acute coronary syndromes.

Authors:  Anita Persson; Marianne Hartford; Johan Herlitz; Thomas Karlsson; Torbjørn Omland; Kenneth Caidahl
Journal:  Heart       Date:  2010-09-28       Impact factor: 5.994

6.  Retrograde blood flow in the aortic arch determined by transesophageal Doppler ultrasound.

Authors:  S Svedlund; R Wetterholm; R Volkmann; K Caidahl
Journal:  Cerebrovasc Dis       Date:  2008-11-15       Impact factor: 2.762

Review 7.  Advances in fetal cardiac imaging.

Authors:  M Sklansky
Journal:  Pediatr Cardiol       Date:  2004 May-Jun       Impact factor: 1.655

8.  Preoperative scallop-by-scallop assessment of mitral prolapse using 2D-transthoracic echocardiography.

Authors:  Giovanni Minardi; Paolo Giuseppe Pino; Carla Clotilde Manzara; Giovanni Pulignano; Giulio Giuseppe Stefanini; Giuseppe Nicola Viceconte; Stefania Leonetti; Andrea Madeo; Carlo Gaudio; Francesco Musumeci
Journal:  Cardiovasc Ultrasound       Date:  2010-01-01       Impact factor: 2.062

9.  Ultrasound evaluation of normal and abnormal fetuses: comparison of conventional, tissue harmonic, and pulse-inversion harmonic imaging techniques.

Authors:  Jeong-Ah Ryu; Bohyun Kim; Sooah Kim; Soon Ha Yang; Moon Hae Choi; Hyeong Sik Ahn
Journal:  Korean J Radiol       Date:  2003 Jul-Sep       Impact factor: 3.500

10.  A realistic flow phantom model of the carotid artery in preterm infants for training and research.

Authors:  Sujith Pereira; Jonathan Reeves; Malcolm Birch; Sakthi Finton-James; Komal Verma; Robert Krug; Ajay Sinha; Stephen Kempley
Journal:  Ultrasound       Date:  2020-02-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.