I Echevarría1, C Barturen, M J Renedo, M C Dios-Viéitez.
Abstract
A sensitive HPLC method with piroxicam as internal standard was developed for assaying amphotericin B in plasma and tissue. An Ultrabase-C18 column and a simple mobile phase consisting of an acetonitrile-acetic acid (10%)-water (41:43:16) mixture were used. The flow-rate was 1 ml/min and the effluent was monitored at 405 nm. The linearity of the assay method was up to 1000 ng/ml and 100 micrograms/g for plasma and tissue, respectively. Intra-day and inter-day RSDs were below 10% for all the sample types. This HPLC assay has been applied to determine amphotericin B in plasma and tissue samples taken during pharmacokinetic and tissue distribution studies in rats.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9781419 DOI: 10.1016/s0021-9673(98)00425-7