| Literature DB >> 977764 |
Abstract
Temperature dependent fluctuations in the erythrocyte sedimentation rate were noted during the assessment of one patient with symptomatic cryoglobulinaemia. Two patients with this condition were found to have erroneously high leucocyte counts when estimated by the Model S Coulter Counter. The artefact was directly related to the degree of exposure to cold before the estimation was performed and appeared to be due to the formation of microaggregates of cryoglobulin and fibrinogen.Entities:
Mesh:
Substances:
Year: 1976 PMID: 977764 PMCID: PMC476209 DOI: 10.1136/jcp.29.10.894
Source DB: PubMed Journal: J Clin Pathol ISSN: 0021-9746 Impact factor: 3.411