Cerebral monitoring by means of oximetry and somatosensory evoked potentials during carotid endarterectomy.

H Cho1, E M Nemoto, H Yonas, J Balzer, R J Sclabassi.   

Abstract

OBJECT: Cerebral ischemia that occurs during carotid endarterectomy is commonly monitored by means of somatosensory evoked potentials (SSEPs) and electroencephalography (EEG). The authors conducted this study to determine whether cerebral ischemia could also be reliably detected by cerebral oximetry.
METHODS: Twenty-nine patients who underwent carotid endarterectomy were monitored by means of SSEPs, EEG, and cerebral oximetry with a model NIRO500 (20 patients) or INVOS3100A (nine patients) oximeter. Changes in amplitude of SSEPs were graded as follows: 0, no change; 1, decrease of less than 50%; 2, decrease of greater than 50%; and 3, 100% decrease. As measured with the NIRO500 oximeter, closing the common caro-tid artery decreased mean oxyhemoglobin levels twice as much (p < 0.005) in the group with SSEPs of 1 to 3 (-13.11+/-5.59 microM [mean+/-standard deviation], 12 patients) as in the group with SSEPs of 0 (-6.22+/-5.59 microM, eight patients). The rise in deoxyhemoglobin was also greater (p < 0.05). Two of nine patients monitored with the INVOS3100A oximeter had SSEPs of 1 and 3, and their regional saturation of oxygen (rSO2) values fell by -11.50 and -11.51, respectively. In the remaining seven patients with SSEPs of 0, the rSO2 ranged between -2.00 and -6.10 with no overlap with the group with SSEPs of I to 3. The increase in oxyhemoglobin monitored using the NIRO500 oximeter and rSO2 monitored using the INVOS3100A machine after opening the external carotid artery was less than that seen after opening the internal carotid artery. Both types of oximeters could detect cerebral ischemia but whereas false negatives occurred with the NIRO500, none was observed with the INVOS3100A. Extracranial contamination was also four times less frequent with the INVOS3100A than with the NIRO500 monitor.
CONCLUSIONS: The results indicate that at least as measured with the INVOS3100A instrument, a decrease in rSO2 of -10 or more or a decrease below an rSO2 of 50 is indicative of cerebral ischemia of sufficient severity to decrease the amplitude of SSEPs.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9761045     DOI: 10.3171/jns.1998.89.4.0533

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  9 in total

1.  The influence of basic ventilation strategies on cerebral oxygenation in anesthetized patients without vascular disease.

Authors:  Paul Picton; Amy Shanks; Perma Dorje; George A Mashour
Journal:  J Clin Monit Comput       Date:  2010-11-09       Impact factor: 2.502

2.  Transcutaneous near-infrared spectroscopy for monitoring spinal cord ischemia: an experimental study in swine.

Authors:  Koichi Suehiro; Tomoharu Funao; Yohei Fujimoto; Akira Mukai; Mitsuyo Nakamura; Kiyonobu Nishikawa
Journal:  J Clin Monit Comput       Date:  2016-08-27       Impact factor: 2.502

3.  Anesthesia for aortic reconstruction in a child with PHACE syndrome.

Authors:  Tatsuyuki Imada; Ryu Okutani; Yutaka Oda
Journal:  J Anesth       Date:  2014-04-18       Impact factor: 2.078

4.  Influence of Ventilation Strategies and Anesthetic Techniques on Regional Cerebral Oximetry in the Beach Chair Position: A Prospective Interventional Study with a Randomized Comparison of Two Anesthetics.

Authors:  Paul Picton; Andrew Dering; Amir Alexander; Mary Neff; Bruce S Miller; Amy Shanks; Michelle Housey; George A Mashour
Journal:  Anesthesiology       Date:  2015-10       Impact factor: 7.892

Review 5.  Systematic review of near-infrared spectroscopy determined cerebral oxygenation during non-cardiac surgery.

Authors:  Henning B Nielsen
Journal:  Front Physiol       Date:  2014-03-17       Impact factor: 4.566

6.  Relationship between intraoperative regional cerebral oxygen saturation trends and cognitive decline after total knee replacement: a post-hoc analysis.

Authors:  Fátima Salazar; Marta Doñate; Teresa Boget; Ana Bogdanovich; Misericordia Basora; Ferran Torres; Isabel Gracia; Neus Fàbregas
Journal:  BMC Anesthesiol       Date:  2014-07-21       Impact factor: 2.217

7.  Non-invasive High Frequency Median Nerve Stimulation Effectively Suppresses Olfactory Intensity Perception in Healthy Males.

Authors:  Ashim Maharjan; Mei Peng; Yusuf O Cakmak
Journal:  Front Hum Neurosci       Date:  2019-01-21       Impact factor: 3.169

8.  Cerebral oxygen saturation after multiple perioperative influential factors predicts the occurrence of postoperative cognitive dysfunction.

Authors:  Cheng Ni; Ting Xu; Nan Li; Yang Tian; Yongzheng Han; Qingsheng Xue; Min Li; Xiangyang Guo
Journal:  BMC Anesthesiol       Date:  2015-10-26       Impact factor: 2.217

9.  Improvement of Olfactory Function With High Frequency Non-invasive Auricular Electrostimulation in Healthy Humans.

Authors:  Ashim Maharjan; Eunice Wang; Mei Peng; Yusuf O Cakmak
Journal:  Front Neurosci       Date:  2018-04-24       Impact factor: 4.677

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.