The use of mass fragmentography in the evaluation of routine methods for glucose determination.

I Björkhem, R Blomstrand, O Falk, G Ohman.   

Abstract

A highly specific and accurate mass fragmentographic reference method for determination of serum glucose is described. A fixed amount of hepta-deuterated glucose is added to a fixed amount of serum. The mixture is lyophilized, converted into the penta-trimethylsilyl-methyloxime derivative and subjected to analysis with a combined gas chromatograph - mass spectrometer equipped with a MID-unit (multiple ion detector). The amount of unlabeled glucose was determined from the ratio between recordings at m/e 319 and 323. The two ions used correspond to the base peak in the mass spectrum of derivative of unlabeled and hepta-deuterium labeled glucose, respectively. The relative standard deviation of the method was 1.9%. The method was compared with different enzymatic methods based on the hexokinase reaction and the glucose oxidase reaction. Of the different methods tested, a glucose oxidase method with determination of maximal rate of consumption of oxygen gave results most close to the results of the reference method.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 975587     DOI: 10.1016/0009-8981(76)90198-4

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  Development and Validation of a Rapid (13)C6-Glucose Isotope Dilution UPLC-MRM Mass Spectrometry Method for Use in Determining System Accuracy and Performance of Blood Glucose Monitoring Devices.

Authors:  Risë K Matsunami; Kimon Angelides; David A Engler
Journal:  J Diabetes Sci Technol       Date:  2015-05-18

2.  Glucose measurement: time for a gold standard.

Authors:  Joakim Hagvik
Journal:  J Diabetes Sci Technol       Date:  2007-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.