Measurement of blood radioactivity for quantification of cerebral blood flow using a gamma camera.

K Murase1, H Fujioka, T Inoue, Y Ishimaru, A Akamune, Y Yamamoto, J Ikezoe.   

Abstract

OBJECTIVE: This study was designed to determine whether gamma cameras can be substituted for well-type scintillation counters in measuring blood radioactivity counts to be used as an input function for the quantitative measurement of cerebral blood flow (CBF).
METHODS: Twelve different aqueous 123I solutions were prepared by serial dilution of the original concentration of 281.9 kBq/ml, and the radioactivity count of each dilution was measured with a gamma camera with the collimator removed, and with a well-type scintillation counter. When measuring the radioactivity counts with a gamma camera, static images were acquired using a 128 x 128 matrix for 5 min, and the regions of interest with 14 x 14 pixels (21 mm x 21 mm) were defined.
RESULTS: There was a good correlation between the results obtained by these two procedures in the range of concentration between 0.008 kBq/ml and 281.9 kBq/ml (y = 4.245x-2.549, r = 1.0, n = 12, s.e.e. = 7.217 kcpm). There was good agreement between the CBF values (ml/100 g/min) obtained using the cross-calibration factor (CCF) and blood radioactivity counts measured with the two procedures (y = 0.990x + 0.552, r = 0.990, n = 231, s.e.e. = 1.340 ml/100 g/min).
CONCLUSION: The results suggest that gamma cameras can be substituted for well-type scintillation counters in the quantitative measurement of CBF, and make it unnecessary to measure CCF after routine calibration of a SPECT apparatus.

Mesh:

Substances:

Year:  1998        PMID: 9755439

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  1 in total

1.  Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator.

Authors:  Toshimune Ito; Yohji Matsusaka; Masahisa Onoguchi; Hajime Ichikawa; Koichi Okuda; Takayuki Shibutani; Masaaki Shishido; Kozo Sato
Journal:  J Appl Clin Med Phys       Date:  2021-01-10       Impact factor: 2.102