Performance of a plastic optical fiber stylet for tracheal intubation of a dog.

D Gravenstein1, S Lampotang, R Melker, R Doviak.   

Abstract

OBJECTIVE: We set out to establish whether a novel plastic optical fiber incorporated into an endotracheal tube (ETT) stylet could be used for intubation of a dog. A secondary objective examined the need for a direct illumination source from a laryngoscope. Lastly, the fragility of the system was tested.
METHODS: An anesthetized dog was repeatedly intubated using a laryngoscope to elevate the tongue and the view of the larynx conducted through the plastic optical fiber stylet (placed within an endotracheal tube) and displayed on a television monitor. Four prototype identical stylets were tested. Repeated intubations were attempted with each stylet and graded as either successful or failed. All four stylets were tested 10 times each using a Miller 4 blade and direct illumination from the laryngoscope. Two of the four stylets were reused during an additional 10 intubation attempts using a Miller 4 blade and laryngoscope (without batteries) with only ambient light. Finally, one stylet was used for intubation after 10, 20, 30, 40 and 50 sharp 90 degree bend-and-straighten cycles using a Miller 4 blade and laryngoscope for direct illumination.
RESULTS: All attempted intubations were successful. However, the image quality was dramatically better when direct illumination from a laryngoscope was used than when ambient light was used. One plastic optical fiber stylet was successfully used to intubate after having been used for 20 intubations and 50 sharp 90 degree bend-and-straighten cycles. A partial lens separation occurred between the 41st and 50th bend cycle but the image remained adequate enough to successfully intubate again.
CONCLUSIONS: A novel plastic optical fiber incorporated into an ETT stylet can be used with a laryngoscope for intubation of a dog. Direct illumination from a laryngoscope provides a better television monitor image than when only ambient light is used. The system was durable, withstanding over 20 uses and 40 sharp bend-and-straighten cycles before a lens separation failure occurred.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9754616     DOI: 10.1023/a:1009961728812

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  4 in total

1.  The broncho-fiberscope as an aid to endotracheal intubation.

Authors:  P A Taylor; R M Towey
Journal:  Br J Anaesth       Date:  1972-06       Impact factor: 9.166

2.  A fibre-optic endoscope used for nasal intubation.

Authors:  P Murphy
Journal:  Anaesthesia       Date:  1967-07       Impact factor: 6.955

3.  A flexible fiber optic laryngoscope.

Authors:  C M Stiles; Q R Stiles; J S Denson
Journal:  JAMA       Date:  1972-09-11       Impact factor: 56.272

4.  The optical stylet--a new intubation technique for adults and children with specific reference to teaching.

Authors:  R L Katz; G Berci
Journal:  Anesthesiology       Date:  1979-09       Impact factor: 7.892

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.