Adaptive lung ventilation (ALV) during anesthesia for pulmonary surgery: automatic response to transitions to and from one-lung ventilation.

N Weiler1, B Eberle, W Heinrichs.   

Abstract

UNLABELLED: Adaptive lung ventilation is a novel closed-loop-controlled ventilation system. Based upon instantaneous breath-to-breath analyses, the ALV controller adjusts ventilation patterns automatically to momentary respiratory mechanics. Its goal is to provide a preset alveolar ventilation (V'A) and, at the same time, minimize the work of breathing. Aims of our study were (1) to investigate changes in respiratory mechanics during transition to and from one-lung ventilation (OLV), (2) to describe the automated adaptation of the ventilatory pattern.
METHODS: With institutional approval and informed consent, 9 patients (33-72 y, 66-88 kg) underwent ALV during total intravenous anesthesia for pulmonary surgery. The ALV controller uses a pressure controlled ventilation mode. V'A is preset by the anesthesiologist. Flow, pressure, and CO2 are continuously measured at the DLT connector. The signals were read into a IBM compatible PC and processed using a linear one-compartment model of the lung to calculate breath-by-breath resistance (R), compliance (C), respiratory time constant (TC), serial dead space (VdS) and V'A. Based upon the results, the controller optimizes respiratory rate (RR) and tidal volume (VT) such as to achieve the preset V'A with the minimum work of breathing. In addition to V'A, only PEEP and FIO2 settings are at the anesthesiologist's discretion. All patients were ventilated using FIO2 = 1,0 and PEEP = 3 cm H2O. Parameters of respiratory mechanics, ventilation, and ABG were recorded during three 5-min periods: 10 min prior to OLV (1), 20 min after onset of OLV (II), and after chest closure (III). Data analyses used nonparametric comparisons of paired samples (Wilcoxon, Friedman) with Bonferroni's correction. Significance was assumed at p < 0.05. Values are given as medians (range).
RESULTS: 20 min after onset of OLV (II), resistance had approximately doubled compared with (1), compliance had decreased from 54 (36-81) to 50 (25-70) ml/cm H2O. TC remained stable at 1.4 (0.8-2.4) vs. 1.2 (0.9)-1.6) s. Institution of OLV was followed by a reproducible response of the ALV controller. The sudden changes in respiratory mechanics caused a transient reduction in VT by 42 (8-59)%, with RR unaffected. In order to reestablish the preset V'A, the controller increased inspiratory pressure in a stepwise fashion from 18 (14-23) to 27 (19-39) cm H2O, thereby increasing VT close to baseline (7.5 (6.6-9.0) ml/kg BW vs. 7.9 (5.4-11.7) ml/kg BW). The controller was, thus, effective in maintaining V'A. The minimum PaO2 during phase II was 101 mmHg. After chest closure, respiratory mechanics had returned to baseline.
CONCLUSIONS: Respiratory mechanics during transition to and from OLV are characterized by marked changes in R and C into opposite directions, leaving TC unaffected. The ALV controller manages these transitions successfully, and maintains V'A reliably without intervention by the anesthesiologist. VT during OLV was found to be consistently lower than recommended in the literature.

Entities:  

Mesh:

Year:  1998        PMID: 9754613     DOI: 10.1023/a:1009974825237

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  17 in total

1.  Mechanics of breathing in man.

Authors:  A B OTIS; W O FENN; H RAHN
Journal:  J Appl Physiol       Date:  1950-05       Impact factor: 3.531

Review 2.  Hypoxic pulmonary vasoconstriction.

Authors:  B E Marshall
Journal:  Acta Anaesthesiol Scand Suppl       Date:  1990

3.  [Adaptive lung ventilation (AVL). Evaluation of new closed loop regulated respiration algorithm for operation in the hyperextended lateral position].

Authors:  N Weiler; B Eberle; F Latorre; S von Paczynski; W Heinrichs
Journal:  Anaesthesist       Date:  1996-10       Impact factor: 1.041

Review 4.  One-lung ventilation and hypoxic pulmonary vasoconstriction: implications for anesthetic management.

Authors:  J L Benumof
Journal:  Anesth Analg       Date:  1985-08       Impact factor: 5.108

5.  Optimization of arterial oxygenation during one-lung anesthesia.

Authors:  L M Capan; H Turndorf; C Patel; S Ramanathan; A Acinapura; J Chalon
Journal:  Anesth Analg       Date:  1980-11       Impact factor: 5.108

6.  An adaptive lung ventilation controller.

Authors:  T P Laubscher; W Heinrichs; N Weiler; G Hartmann; J X Brunner
Journal:  IEEE Trans Biomed Eng       Date:  1994-01       Impact factor: 4.538

7.  Positive end-expiratory pressure during one-lung ventilation improves oxygenation in patients with low arterial oxygen tensions.

Authors:  E Cohen; J B Eisenkraft
Journal:  J Cardiothorac Vasc Anesth       Date:  1996-08       Impact factor: 2.628

8.  The AVL-mode: a safe closed loop algorithm for ventilation during total intravenous anesthesia.

Authors:  N Weiler; W Heinrichs; W Kessler
Journal:  Int J Clin Monit Comput       Date:  1994-05

9.  Continuous spirometry for detection of double-lumen endobronchial tube displacement.

Authors:  G I Bardoczky; M Levarlet; E Engelman; P deFrancquen
Journal:  Br J Anaesth       Date:  1993-05       Impact factor: 9.166

10.  Respiratory mechanics by least squares fitting in mechanically ventilated patients: applications during paralysis and during pressure support ventilation.

Authors:  G A Iotti; A Braschi; J X Brunner; T Smits; M Olivei; A Palo; R Veronesi
Journal:  Intensive Care Med       Date:  1995-05       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.