M F Schreuder1, A P van Driel, A van Lingen, J C Roos, C M de Ridder, R A Manoliu, E F David, J C Netelenbos.
Abstract
Measurements of bone mineral density using dual-energy X-ray absorptiometry (DXA) gives area values (g cm-2) rather than true volumetric values (g cm-3). To calculate the vertebral volume using planar postero-anterior and lateral DXA values, several different geometrical approximations were used: cubic, cylindrical with a circular cross-section and cylindrical with an elliptical cross-section. The aim of this study was to compare these geometrical approximations with each other and with a reference standard, defined as the volume found on a computed tomographic (CT) scan. L2 and L3 were evaluated in a phantom study. Volume approximations by the cube or cylinder with circular cross-section geometry showed more than a 50% overestimation (range 54-74%). However, the elliptical cylinder approach showed very good agreement: 2.1% and 1.2% for L2 and L3, respectively, when compared to the CT volumes. In addition, we performed four patient studies with both CT and DXA to evaluate the elliptical cylinder estimate in a clinical setting. For L2 and L3, the mean relative difference was less than 2%. We conclude that the elliptical cylinder approach results in the most accurate bone volume estimates in both the phantom and patients.Entities:
Mesh:
Year: 1998 PMID: 9751926 DOI: 10.1097/00006231-199808000-00002
Source DB: PubMed Journal: Nucl Med Commun ISSN: 0143-3636 Impact factor: 1.690