Literature DB >> 9745937

Intraluminal signal intensity of iliac artery stents investigated by contrast-enhanced three-dimensional MR angiography.

Y Amano1, K Gemma, H Kawamata, T Kumazaki.   

Abstract

The purpose of this study is to investigate the intraluminal signal of iliac artery stents by contrast-enhanced three-dimensional magnetic resonance (MR) angiography with a short echo time. Ten patients with iliac arterial diseases treated by intravascular metallic stent placement were examined with a 1.5 T MR imager. Gadolinium-enhanced MR angiography depicted the intraluminal signal in Strecker stent, but not in the Wallstent and Palmaz stent. Strecker stent made of nonferromagnetic tantalum was the best-suited stent for use with contrast-enhanced MR angiography.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9745937     DOI: 10.1016/s0895-6111(97)00040-2

Source DB:  PubMed          Journal:  Comput Med Imaging Graph        ISSN: 0895-6111            Impact factor:   4.790


  1 in total

1.  CT and MR imaging findings and their implications in the follow-up of patients with intracranial aneurysms treated with endosaccular occlusion with onyx.

Authors:  Isil Saatci; H Saruhan Cekirge; Elisa F M Ciceri; Michel E Mawad; A Gulsun Pamuk; Aytekin Besim
Journal:  AJNR Am J Neuroradiol       Date:  2003-04       Impact factor: 3.825