Development of phantoms for spiral CT.

D J Goodenough1, J R Levy, C Kasales.   

Abstract

PURPOSE: This paper reports on the development of a new phantom for spiral CT. The phantom meets the increased demands on phantom z-axis uniformity in order that objects from the CT slice, immediately above and below the CT slice of interest, do not contribute perturbing information to the reconstructed CT slice.
MATERIAL AND METHODS: The phantom depends on formulation of tissue-like materials that can be cast and produced in both geometric and anthropomorphic shapes with sufficient z-axis length to enable unperturbed CT slices of test objects of interest. These materials are then used to produce a series of test objects of CT image quality including low contrast samples that do not require volume averaging or mixing of solutions, and that can reflect sub-slice thickness test objects and supra-slice thickness test objects.
RESULTS: The overall phantom and its individual test objects provides meaningful tests of spiral CT image quality including slice sensitivity, CT number linearity and tests of high and low contrast resolution. Schematic designs and actual CT scans are shown.
CONCLUSION: The new spiral phantom appears to meet the increased demands of spiral CT on phantom design, particularly z-axis length, and requirements for low contrast resolution test objects.

Mesh:

Substances:

Year:  1998        PMID: 9740042     DOI: 10.1016/s0895-6111(98)00020-2

Source DB:  PubMed          Journal:  Comput Med Imaging Graph        ISSN: 0895-6111            Impact factor:   4.790


  1 in total

1.  Method and phantom to study combined effects of in-plane (x,y) and z-axis resolution for 3D CT imaging.

Authors:  David Goodenough; Josh Levy; Smari Kristinsson; Jesper Fredriksson; Hildur Olafsdottir; Austin Healy
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.