Literature DB >> 9738544

In situ temperature measurements with thermocouple probes during laser interstitial thermotherapy (LITT): quantification and correction of a measurement artifact.

F Manns1, P J Milne, X Gonzalez-Cirre, D B Denham, J M Parel, D S Robinson.   

Abstract

BACKGROUND AND
OBJECTIVE: The purpose of this work was to quantify the magnitude of an artifact induced by stainless steel thermocouple probes in temperature measurements made in situ during experimental laser interstitial thermo-therapy (LITT). A procedure for correction of this observational error is outlined. STUDY DESIGN/
MATERIALS AND METHODS: A CW Nd:YAG laser system emitting 20W for 25-30 s delivered through a fiber-optic probe was used to create localized heating. The temperature field around the fiber-optic probe during laser irradiation was measured every 0.3 s in air, water, 0.4% intralipid solution, and fatty cadaver pig tissue, with a field of up to fifteen needle thermocouple probes.
RESULTS: Direct absorption of Nd:YAG laser radiation by the thermocouple probes induced an overestimation of the temperature, ranging from 1.8 degrees C to 118.6 degrees C in air, 2.2 degrees C to 9.9 degrees C in water, 0.7 C to 4.7 C in intralipid and 0.3 C to 17.9 C in porcine tissue after irradiation at 20W for 30 s and depending on the thermocouple location. The artifact in porcine tissue was removed by applying exponential and linear fits to the measured temperature curves.
CONCLUSION: Light absorption by thermocouple probes can induce a significant artifact in the measurement of laser-induced temperature increases. When the time constant of the thermocouple effect is much smaller than the thermal relaxation time of the surrounding tissue, the artifact can be accurately quantified. During LITT experiments where temperature differences of a few degrees are significant, the thermocouple artifact must be removed in order to be able accurately to predict the treatment outcome.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9738544     DOI: 10.1002/(sici)1096-9101(1998)23:2<94::aid-lsm7>3.0.co;2-q

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  13 in total

1.  The influence of Nd:YAG laser irradiation on Fluoroptic® temperature measurement: an experimental evaluation.

Authors:  Frank Hübner; Babak Bazrafshan; Jörg Roland; Antje Kickhefel; Thomas Josef Vogl
Journal:  Lasers Med Sci       Date:  2012-04-04       Impact factor: 3.161

2.  Temperature elevation profile inside the rat brain induced by a laser beam.

Authors:  Ali Ersen; Ammar Abdo; Mesut Sahin
Journal:  J Biomed Opt       Date:  2014-01       Impact factor: 3.170

3.  A comparison of temperature profile depending on skin types for laser hair removal therapy.

Authors:  Tae-Hoon Kim; Gwi-Won Lee; Jong-In Youn
Journal:  Lasers Med Sci       Date:  2014-05-15       Impact factor: 3.161

4.  Residual tumor after laser ablation of human non-small-cell lung cancer demonstrated by ex vivo staining: correlation with invasive temperature measurements.

Authors:  Christian Oliver Martin Hoffmann; Christian Rosenberg; Albert Linder; Norbert Hosten
Journal:  MAGMA       Date:  2011-06-09       Impact factor: 2.310

Review 5.  Image-guided thermal ablation with MR-based thermometry.

Authors:  Mingming Zhu; Ziqi Sun; Chin K Ng
Journal:  Quant Imaging Med Surg       Date:  2017-06

Review 6.  Laser Ablation for Cancer: Past, Present and Future.

Authors:  Emiliano Schena; Paola Saccomandi; Yuman Fong
Journal:  J Funct Biomater       Date:  2017-06-14

7.  Photothermal Response of Polyhydroxy Fullerenes.

Authors:  Alan Chen; Stephen R Grobmyer; Vijay B Krishna
Journal:  ACS Omega       Date:  2020-06-04

8.  Photoacoustic temperature imaging based on multi-wavelength excitation.

Authors:  Lei Meng; Olivier Deschaume; Lionel Larbanoix; Eduard Fron; Carmen Bartic; Sophie Laurent; Mark Van der Auweraer; Christ Glorieux
Journal:  Photoacoustics       Date:  2018-11-22

Review 9.  Optical fiber-based MR-compatible sensors for medical applications: an overview.

Authors:  Fabrizio Taffoni; Domenico Formica; Paola Saccomandi; Giovanni Di Pino; Emiliano Schena
Journal:  Sensors (Basel)       Date:  2013-10-18       Impact factor: 3.576

Review 10.  Fiber Optic Sensors for Temperature Monitoring during Thermal Treatments: An Overview.

Authors:  Emiliano Schena; Daniele Tosi; Paola Saccomandi; Elfed Lewis; Taesung Kim
Journal:  Sensors (Basel)       Date:  2016-07-22       Impact factor: 3.576