R A Potyrailo1, R C Conrad, A D Ellington, G M Hieftje.
Abstract
A flexible biosensor has been developed that utilizes immobilized nucleic acid aptamers to specifically detect free nonlabeled non-nucleic acid targets such as proteins. In a model system, an anti-thrombin DNA aptamer was fluorescently labeled and covalently attached to a glass support. Thrombin in solution was selectively detected by following changes in the evanescent-wave-induced fluorescence anisotropy of the immobilized aptamer. The new biosensor can detect as little as 0.7 amol of thrombin in a 140-pL interrogated volume, has a dynamic range of 3 orders of magnitude, has an inter-sensing-element measurement precision of better than 4% RSD over the range 0-200 nM, and requires less than 10 min for sample analysis. The aptamer-sensor format is generalizable and should allow sensitive, selective, and fast determination of a wide range of analytes.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9726167 DOI: 10.1021/ac9802325
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986