Verification of the omni wedge technique.

B D Milliken1, J V Turian, R J Hamilton, S J Rubin, F T Kuchnir, C X Yu, J W Wong.   

Abstract

The optimal field shape achieved using a multileaf collimator (MLC) often requires collimator rotation to minimize the adverse effects of the scalloped dose distribution the leaf steps produce. However, treatment machines are designed to deliver wedged fields parallel or perpendicular to the direction of the leaves. An analysis of cases from our clinic showed that for 25% of the wedged fields used to treat brain and lung tumors, the wedge direction and optimal MLC orientation differed by 20 degrees or more. The recently published omni wedge technique provides the capability of producing a wedged field with orientation independent of the orientation of the collimator. This paper presents a comparison of the three-dimensional (3D) dose distributions of the omni wedged field with distributions of wedged fields produced using both the universal and dynamic wedge techniques. All measurements were performed using film dosimetry techniques. The omni wedge generated fields closely matched the conventional wedged fields. Throughout 95% of the irradiated volume (excluding the penubra), the dose distribution of the omni wedged field ranged from +5.5 to -3.5 +/- 1.5% of that of the conventionally wedged fields. Calculation of the omni wedged field is as accurate as conventional wedged field calculation when using a 3D treatment planning systems. For two-dimensional treatment planning systems, where one must assume that the omni wedged field is identical to a conventional field, the calculated field and the delivered field differs by a small amount.

Entities:  

Mesh:

Year:  1998        PMID: 9725128     DOI: 10.1118/1.598314

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Comparison of two algorithms for determining beam weights and wedge filters.

Authors:  Jianrong Dai; Yunping Zhu
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.