Literature DB >> 9722845

MR imaging of the inner ear: comparison of a three-dimensional fast spin-echo sequence with use of a dedicated quadrature-surface coil with a gadolinium-enhanced spoiled gradient-recalled sequence.

S Naganawa1, T Ito, H Fukatsu, T Ishigaki, T Nakashima, N Ichinose, Y Kassai, M Miyazaki.   

Abstract

PURPOSE: To prospectively evaluate the sensitivity and specificity of magnetic resonance (MR) imaging in the inner ear with a long echo train, three-dimensional (3D), asymmetric Fourier-transform, fast spin-echo (SE) sequence with use of a dedicated quadrature-surface phased-array coil to detect vestibular schwannoma in the cerebellopontine angle and the internal auditory canal.
MATERIALS AND METHODS: In 205 patients (410 ears) with ear symptoms, 1.5-T MR imaging was performed with unenhanced 3D asymmetric fast SE and gadolinium-enhanced 3D gradient-recalled (SPGR) sequences with use of a quadrature surface phased-array coil. The 3D asymmetric fast SE images were reviewed by two radiologists, with the gadolinium-enhanced 3D SPGR images used as the standard of reference.
RESULTS: Nineteen lesions were detected in the 410 ears (diameter range, 2-30 mm; mean, 10.5 mm +/- 6.4 [standard deviation]; five lesions were smaller than 5 mm). With 3D asymmetric fast SE, sensitivity, specificity, and accuracy, respectively, were 100%, 99.5%, and 99.5% for observer 1 and 100%, 99.7%, and 99.8% for observer 2.
CONCLUSION: The unenhanced 3D asymmetric fast SE sequence with a quadrature-surface phased-array coli allows the reliable detection of vestibular schwannoma in the cerebellopontine angle and internal auditory canal.

Entities:  

Mesh:

Year:  1998        PMID: 9722845     DOI: 10.1148/radiology.208.3.9722845

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  9 in total

1.  MR imaging of the enlarged endolymphatic duct and sac syndrome by use of a 3D fast asymmetric spin-echo sequence: volume and signal-intensity measurement of the endolymphatic duct and sac and area measurement of the cochlear modiolus.

Authors:  S Naganawa; T Koshikawa; E Iwayama; H Fukatsu; T Ishiguchi; T Ishigaki; M Ikeda; T Nakashima; N Ichinose
Journal:  AJNR Am J Neuroradiol       Date:  2000-10       Impact factor: 3.825

2.  MR cisternography of the cerebellopontine angle: comparison of three-dimensional fast asymmetrical spin-echo and three-dimensional constructive interference in the steady-state sequences.

Authors:  S Naganawa; T Koshikawa; H Fukatsu; T Ishigaki; T Fukuta
Journal:  AJNR Am J Neuroradiol       Date:  2001 Jun-Jul       Impact factor: 3.825

3.  Fast recovery 3D fast spin-echo MR imaging of the inner ear at 3 T.

Authors:  Shinji Naganawa; Tokiko Koshikawa; Hiroshi Fukatsu; Takeo Ishigaki; Ikuo Aoki; Ayako Ninomiya
Journal:  AJNR Am J Neuroradiol       Date:  2002-02       Impact factor: 3.825

4.  High-resolution T1-weighted 3D real IR imaging of the temporal bone using triple-dose contrast material.

Authors:  Shinji Naganawa; Tokiko Koshikawa; Tatsuya Nakamura; Hiroshi Fukatsu; Takeo Ishigaki; Ikuo Aoki
Journal:  Eur Radiol       Date:  2003-05-07       Impact factor: 5.315

5.  Diagnostic accuracy of screening MR imaging using unenhanced axial CISS and coronal T2WI for detection of small internal auditory canal lesions.

Authors:  T A Abele; D A Besachio; E P Quigley; R K Gurgel; C Shelton; H R Harnsberger; R H Wiggins
Journal:  AJNR Am J Neuroradiol       Date:  2014-07-17       Impact factor: 3.825

6.  Accelerated Internal Auditory Canal Screening Magnetic Resonance Imaging Protocol With Compressed Sensing 3-Dimensional T2-Weighted Sequence.

Authors:  Mikell Yuhasz; Michael J Hoch; Mari Hagiwara; Mary T Bruno; James S Babb; Esther Raithel; Christoph Forman; Abbas Anwar; J Thomas Roland; Timothy M Shepherd
Journal:  Invest Radiol       Date:  2018-12       Impact factor: 6.016

7.  High-spatial-resolution MR cisternography of the cerebellopontine angle in 90 seconds with a zero-fill interpolated fast recovery 3D fast asymmetric spin-echo sequence.

Authors:  Tatsuya Nakamura; Shinji Naganawa; Tokiko Koshikawa; Hiroshi Fukatsu; Yasuo Sakurai; Ikuo Aoki; Ayako Ninomiya; Takeo Ishigaki
Journal:  AJNR Am J Neuroradiol       Date:  2002-09       Impact factor: 3.825

8.  Flow ghost artifact in slice-encoding direction mimicking vestibular schwannoma in contrast-enhanced 3D spoiled gradient-echo sequence.

Authors:  Shinji Naganawa; Tokiko Koshikawa; Hiroshi Fukatsu; Takeo Ishigaki; Ikuo Aoki
Journal:  Eur Radiol       Date:  2003-11-05       Impact factor: 5.315

9.  MR imaging of the internal auditory canal and inner ear at 3T: comparison between 3D driven equilibrium and 3D balanced fast field echo sequences.

Authors:  Jun Soo Byun; Hyung Jin Kim; Yoo Jeong Yim; Sung Tae Kim; Pyoung Jeon; Keon Ha Kim; Sam Soo Kim; Yong Hwan Jeon; Jiwon Lee
Journal:  Korean J Radiol       Date:  2008 May-Jun       Impact factor: 3.500

  9 in total