Literature DB >> 9720937

Validation of a new transit time ultrasound flowmeter for measuring blood flow in colonic mesenteric arteries.

F V Mortensen1, J S Rasmussen, O Viborg, S Laurberg, E M Pedersen.   

Abstract

OBJECTIVE: To validate a transit time ultrasound flowmeter (CardioMed CM 4000) for measuring blood flow in isolated colonic mesenteric arteries.
DESIGN: Experimental and clinical study.
SETTING: Teaching hospital, Denmark. ANIMALS AND
SUBJECTS: One female pig, and 6 patients being operated on for carcinoma of the sigmoid colon and rectum.
INTERVENTIONS: Volume blood flow measured by Cardiomed CM 4000 and pump withdrawal flow recordings. MAIN OUTCOME MEASURES: Correlation between the two methods.
RESULTS: There was good agreement between transit time flow recordings and pump withdrawal flow recordings (correlation coefficient of 1.0). Of the differences between the two methods, 95% were between -0.16 ml min(-1) and 1.29 ml min(-1), mean 0.57 ml min(-1), or (in percentages) 105, 95% lying between 97-115. There was also good reproducibility in transit time flow recordings, the mean difference between repeated measurements being 0.06 ml min(-1), 95% lying between -0.66 ml min(-1) and 0.78 ml min(-1).
CONCLUSION: Ultrasound transit time flow recordings gave precise measurements of blood flow in isolated colonic mesenteric arteries.

Entities:  

Mesh:

Year:  1998        PMID: 9720937     DOI: 10.1080/110241598750005705

Source DB:  PubMed          Journal:  Eur J Surg        ISSN: 1102-4151


  3 in total

1.  Hepatic blood perfusion measured by 3-minute dynamic 18F-FDG PET in pigs.

Authors:  Michael Winterdahl; Ole Lajord Munk; Michael Sørensen; Frank Viborg Mortensen; Susanne Keiding
Journal:  J Nucl Med       Date:  2011-06-16       Impact factor: 10.057

2.  Tracer input for kinetic modelling of liver physiology determined without sampling portal venous blood in pigs.

Authors:  Michael Winterdahl; Susanne Keiding; Michael Sørensen; Frank Viborg Mortensen; Aage Kristian Olsen Alstrup; Ole Lajord Munk
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-09-30       Impact factor: 9.236

3.  Hepatic blood perfusion estimated by dynamic contrast-enhanced computed tomography in pigs: limitations of the slope method.

Authors:  Michael Winterdahl; Michael Sørensen; Susanne Keiding; Frank V Mortensen; Aage K O Alstrup; Søren B Hansen; Ole L Munk
Journal:  Invest Radiol       Date:  2012-10       Impact factor: 6.016

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.