| Literature DB >> 9719332 |
G Maleux1, H Rousseau, P Otal, D Colombier, Y Glock, F Joffre.
Abstract
Transluminal placement of a stent graft in patients with an abdominal aortic aneurysm is a new endovascular technique that offers a potentially less invasive and less risky alternative to open surgery. Complications after stent graft placement are not infrequent, but in most cases secondary endovascular intervention is successful. We describe a late major leak in the aneurysmal sac caused by a distal migration of the iliac limb of a bifurcated graft. This late complication was successfully treated by covered stent placement, excluding and thrombosing completely the reformed aneurysm.Entities:
Mesh:
Year: 1998 PMID: 9719332 DOI: 10.1016/s0741-5214(98)70172-7
Source DB: PubMed Journal: J Vasc Surg ISSN: 0741-5214 Impact factor: 4.268