Confirmation of epidural catheter placement using nerve stimulation.

B C Tsui1, S Gupta, B Finucane.   

Abstract

PURPOSE: To examine the reliability of low current electrical epidural stimulation to confirm epidural catheter placement.
METHODS: Forty patients with epidural catheters (19G Arrow Flextip plus) already in place for post-operative pain management were studied. An adapter (Arrow-Johans ECG Adapter) was attached to the connector of the epidural catheter. The epidural catheter and adapter were filled with normal saline. The cathode lead of the nerve stimulator was attached to the metal hub of the adapter. Catheter placement was judged to be correct or incorrect, depending upon the presence or absence of truncal or limb movement to 1 Hz stimulation (1-10 mA). A standard test dose (3 ml lidocaine 1.5% with 1:200,000 epinephrine) was then injected. The efficacy of the epidural morphine was assessed independently.
RESULTS: The sensitivity and specificity of the test was 100% and 91.6% compared with the standard test dose. The positive and negative predictive value was 96% and 100%. In predicting the clinical effect of epidural morphine, the sensitivity and specificity was 96.1% and 76.9%. The positive and negative predictive value was 89% and 90%. The correlation of unilateral or bilateral motor response from the test and sensory response from the lidocaine test with sensitivity and specificity was 91.6% and 53.0%. The predictive value for unilateral response was 61% and for bilateral was 88%.
CONCLUSION: This study establishes this test as a simple, objective and reliable technique for confirmation of epidural catheter placement.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9717595     DOI: 10.1007/BF03012093

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  14 in total

1.  A plantar flexion response to nerve stimulation indicates needle misplacement in the epidural/spinal space during psoas compartment block.

Authors:  Mehmet Cesur; Haci A Alici; Ali F Erdem
Journal:  J Anesth       Date:  2009-02-22       Impact factor: 2.078

2.  Lateralizing epidural catheter placement as evidenced by electrical stimulation.

Authors:  Nigel E Sharrock; George Go
Journal:  Clin Orthop Relat Res       Date:  2014-05       Impact factor: 4.176

3.  False negative electrical epidural stimulation test due to the excessive impedance of a saline bridge.

Authors:  Jordan D Hill; Ban C H Tsui
Journal:  Can J Anaesth       Date:  2021-01-25       Impact factor: 5.063

Review 4.  Innovative approaches to neuraxial blockade in children: the introduction of epidural nerve root stimulation and ultrasound guidance for epidural catheter placement.

Authors:  Ban C H Tsui
Journal:  Pain Res Manag       Date:  2006       Impact factor: 3.037

5.  Epidural catheter malposition in a failed epidural anesthesia confirmed by computed tomography.

Authors:  Se Jin Lee; Sang Ho Kim; Sun Young Park; Mun Gyu Kim; Bo Il Jung; Si Young Ok
Journal:  Korean J Pain       Date:  2011-02-25

6.  Strategic placement of epidural catheter.

Authors:  Jae-Hang Shim
Journal:  Korean J Anesthesiol       Date:  2012-09-14

Review 7.  Localization of epidural space: A review of available technologies.

Authors:  Hesham Elsharkawy; Abraham Sonny; Ki Jinn Chin
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2017 Jan-Mar

8.  Electric stimulation-guided epidural analgesia for vaginal delivery: A randomized prospective study.

Authors:  Chung Hun Lee; Sang Sik Choi; Mi Kyoung Lee; Jung Eun Kim; Dong Ik Chung; Mido Lee
Journal:  PLoS One       Date:  2019-01-11       Impact factor: 3.240

9.  Efficacy of electrical stimulation on epidural anesthesia for cesarean section: a randomized controlled trial.

Authors:  Young Sung Kim; Hyo Sung Kim; Hyerim Jeong; Chung Hun Lee; Mi Kyoung Lee; Sang Sik Choi
Journal:  BMC Anesthesiol       Date:  2020-06-10       Impact factor: 2.217

10.  Minimum current requirement for confirming the localization of an epiradicular catheter placement.

Authors:  Ji Seon Jeong; Jae Chol Shim; Jae Hang Shim; Dong Won Kim; Min Serk Kang
Journal:  Korean J Anesthesiol       Date:  2012-09-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.