G P Liney1, L W Turnbull, A J Knowles.
Abstract
A simple method is described that corrects for the intensity inhomogeneity experienced when using endorectal surface coils in prostate imaging. The method uses a series of proton-density-weighted images, which are acquired to demonstrate the sensitivity profile of the coil and used to correct T2-weighted images acquired in the same plane. The results demonstrate improved image uniformity and visualization of prostatic anatomy.Mesh:
Year: 1998 PMID: 9702904 DOI: 10.1002/jmri.1880080432
Source DB: PubMed Journal: J Magn Reson Imaging ISSN: 1053-1807 Impact factor: 4.813