| Literature DB >> 9702716 |
E W Hsu1, S B Reeder, J R MacFall.
Abstract
Adiabatic pulses, although useful in generating uniform spin nutation in the presence of inhomogeneous B1 fields, are limited for NMR imaging applications due to the lack of slice-selective excitation capability. Selective excitation techniques using gradient modulation have been introduced; however, present methods require either a minimum of two excitations or eight adiabatic segments. Here, a scheme is presented that allows single-shot, arbitrary flip-angle, and slice-selective excitation with only four adiabatic half-passage segments. The technique is demonstrated via computer simulation and experimental tests on a phantom. Furthermore, issues associated with the implementation of these gradient-modulated adiabatic pulses are discussed.Mesh:
Year: 1998 PMID: 9702716 DOI: 10.1002/mrm.1910400220
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668