A sensitive and specific high-performance liquid chromatographic method was developed and validated for the determination of minocycline in human plasma. The method uses liquid-liquid extraction, reextraction and HPLC with UV detection. The assay shows linearity in the tested range of 28-3533 ng/ml with a limit of quantification of 30 ng/ml. The inter-day precision was found to be +/-3.84 to +/-6.57% (R.S.D.) in the range of 148 to 2743 ng/ml. The method was successfully applied to pharmacokinetic studies.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9691330 DOI: 10.1016/s0021-9673(98)00402-6
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759