Correction of pressure waveforms recorded by fluid-filled catheter recording systems: a new method using a transfer equation.

B Lambermont1, P Gerard, O Detry, P Kolh, P Potty, V D'orio, R Marcelle.   

Abstract

BACKGROUND: Pressure measuring systems using fluid-filled catheters can result in the recording of distorted pressure waveforms. It results in phase delay, overestimation of systolic and, to a lesser extent, of diastolic pressure. We designed and evaluated a method to correct this pressure waveform distortion using an appropriate transfer equation obtained from the dynamic response of the fluid-filled catheter. This transfer equation is based on the principle that a fluid-filled catheter recording system is considered as an underdamped dynamic system fully characterized by its natural frequency (omega n) and damping ratio (zeta).
METHODS: Pressure waveforms, simultaneously recorded in vitro or in vivo by a fluid-filled catheter (Pc) and a micromanometer-tipped catheter (Pref), were used to validate the method. Dynamic response of the catheter used was obtained from a fastflush test. The corrected signal (Ppred) was obtained using omega n, zeta and the following transfer equation: d2Pc/dt2 + 2 omega n zeta dPc/dt + omega n 2Pc = C Ppred (t) After correction of Pc, Ppred was compared, using a linear regression, with Pref taken as reference.
RESULTS: Our results showed that Ppred was fitted to Pref with excellent coefficient correlation (0.99). The mean error and the standard error of estimate were respectively -1.16 mmHg and 1.4 mmHg.
CONCLUSION: This new method can convert the distorted pressure waveforms transmitted by any fluid-filled catheters into high-fidelity signals. It suppresses the phase delay and the over-estimation of systolic pressure induced by fluid-filled catheters.

Mesh:

Year:  1998        PMID: 9689280     DOI: 10.1111/j.1399-6576.1998.tb05307.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  3 in total

1.  Low cost circulatory pressure acquisition and fluid infusion rate measurement system for clinical research.

Authors:  Rachel Smith; Amelia Rolfe; Chris Cameron; Geoffrey M Shaw; J Geoffrey Chase; Christopher G Pretty
Journal:  HardwareX       Date:  2022-05-21

2.  Arterial dP/dtmax accurately reflects left ventricular contractility during shock when adequate vascular filling is achieved.

Authors:  Philippe Morimont; Bernard Lambermont; Thomas Desaive; Nathalie Janssen; Geoffrey Chase; Vincent D'Orio
Journal:  BMC Cardiovasc Disord       Date:  2012-03-01       Impact factor: 2.298

3.  Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study.

Authors:  Stefano Romagnoli; Zaccaria Ricci; Diego Quattrone; Lorenzo Tofani; Omar Tujjar; Gianluca Villa; Salvatore M Romano; A Raffaele De Gaudio
Journal:  Crit Care       Date:  2014-11-30       Impact factor: 9.097

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.