Literature DB >> 9684717

Influence of body position on defibrillation thresholds of nonthoracotomy implantable defibrillators: a prospective randomized evaluation.

P Schauerte1, B Diem, K Ziegert, A Franke, P Hanrath, C Stellbrink.   

Abstract

INTRODUCTION: Defibrillation thresholds (DFTs) usually are determined with the patient in the supine position. However, patients may be in the upright position when a shock is delivered during follow-up, which may explain some first shock failures observed clinically. This study investigated whether body posture affects defibrillation energy requirements of nonthoracotomy implantable cardioverter defibrillators with biphasic shocks. METHODS AND
RESULTS: Using a step up-down protocol, DFTs were compared intraindividually in 52 patients ("active-can" sytems in 41 patients, two-lead systems in 11 patients) for the supine and upright positions as achieved by a tilt table. The mean DFT was 7.3 +/- 4.2 J in the supine versus 9.2 +/- 4.8 J in the upright position (P = 0.002). Repeated comparison in reversed order 3 months after implantation in 22 patients revealed thresholds of 6.2 +/- 2.5 J (supine) versus 8.4 +/- 3.7 J (upright; P < 0.03) 1 week and 4.4 +/- 2.4 J (supine) versus 6.2 +/- 4.1 J (upright; P < 0.04) 3 months after implantation. DFTs decreased significantly for both body positions from 1 week to 3 months after implantation (P < 0.04).
CONCLUSION: (1) DFTs for biphasic shocks delivered by nonthoracotomy defibrillators are higher in the upright compared to the supine body position. (2) Differences remain significant 3 months after implantation. For both body positions, DFT decreases significantly from 1 week to 3 months after implantation. These findings have important implications for programming first shock energy to lower than maximal values or for development of devices with lower maximal stored energy.

Entities:  

Mesh:

Year:  1998        PMID: 9684717     DOI: 10.1111/j.1540-8167.1998.tb00956.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  1 in total

1.  Orthostatic increase in defibrillation threshold leading to defibrillation failure and prolonged cardiac arrest in a sitting position: Lessons from a patient's near-fatal experience.

Authors:  Xiaoke Liu; Samuel Asirvatham; G Glenn Kabell
Journal:  HeartRhythm Case Rep       Date:  2016-03-28