Literature DB >> 9683107

Evaluation of an automated microbiologic blood culture device for detection of bacteria in platelet components.

S J Wagner1, D Robinette.   

Abstract

BACKGROUND: Automated culture methods have been used by several investigators to detect bacterial contamination of cellular blood components. We investigated several factors affecting detection by automated culture of bacteria in platelet concentrates (PCs). These factors included the initial contamination level in PCs, the PC sample volume, the PC sample time, and the white cell level in relation to bacteria levels in the PCs. STUDY DESIGN AND METHODS: Staphylococcus epidermidis or Escherichia coli was inoculated into freshly prepared PCs or white cell-reduced PCs to yield colony-forming unit (CFU) levels of 10, 1, or 0.1 per mL. At the time of inoculation (t=0) and at t=6, t=24, and t=48 hours, 0.5, 1.0, and 2.0 mL samples of the contaminated PCs were transferred into culture bottles. The presence of bacteria in the culture bottles was subsequently monitored by an automated blood culturing instrument. Bacteria levels in the PC at the time of first automated culture detection were determined by quantitative plating.
RESULTS: E. coli was detected in 92 percent of experiments when 1.0- or 2.0-mL samples were taken at t=6 hours. At t=24 hours, 100-percent detection was observed with all tested inoculation volumes; however, by that time, >10(7) CFU per mL of bacteria were present in every PC. For S. epidermidis, 89 percent and 83 percent of contaminated PCs were detected with a t=24 hour sampling time and 2.0- or 1.0-mL sampling volume. Seven of 36 PCs with a 2.0-mL sampling volume and 10 of 36 PCs with a 1.0-mL sampling volume contained >10(6) CFU per mL of S. epidermidis at the time of first detection.
CONCLUSION: Data from this preliminary evaluation suggest that sampling times of 24 hours or more would be necessary to provide confidence in detection of E. coli or S. epidermidis in PCs using this culture method.

Entities:  

Mesh:

Year:  1998        PMID: 9683107     DOI: 10.1046/j.1537-2995.1998.38798346637.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  5 in total

1.  Effects of storage temperature on hematopoietic stability and microbial safety of BM aspirates.

Authors:  S Hahn; W Sireis; K Hourfar; D Karpova; K Dauber; V A J Kempf; E Seifried; M Schmidt; H Bönig
Journal:  Bone Marrow Transplant       Date:  2013-11-04       Impact factor: 5.483

2.  Two novel real-time reverse transcriptase PCR assays for rapid detection of bacterial contamination in platelet concentrates.

Authors:  Jens Dreier; Melanie Störmer; Knut Kleesiek
Journal:  J Clin Microbiol       Date:  2004-10       Impact factor: 5.948

3.  Rapid screening method for detection of bacteria in platelet concentrates.

Authors:  S Ribault; K Harper; L Grave; C Lafontaine; P Nannini; A Raimondo; I Besson Faure
Journal:  J Clin Microbiol       Date:  2004-05       Impact factor: 5.948

4.  Gum arabic modified Fe3O4 nanoparticles cross linked with collagen for isolation of bacteria.

Authors:  Ashwin Murugappan Chockalingam; Heman Kumar Ramiya Ramesh Babu; Raghuraman Chittor; Jai Prakash Tiwari
Journal:  J Nanobiotechnology       Date:  2010-12-15       Impact factor: 10.435

5.  Cationized Magnetoferritin Enables Rapid Labeling and Concentration of Gram-Positive and Gram-Negative Bacteria in Magnetic Cell Separation Columns.

Authors:  S Correia Carreira; J Spencer; W Schwarzacher; A M Seddon
Journal:  Appl Environ Microbiol       Date:  2016-05-31       Impact factor: 4.792