An empirical relationship for determining photon beam quality in TG-21 from a ratio of percent depth doses.

D S Followill1, R C Tailor, V M Tello, W F Hanson.   

Abstract

A key component of the Radiological Physics Center's (RPC) on-site dosimetry review visits are photon beam calibrations for which determination of the energy of the x ray is a key element. The ratio of ionizations, TPR20/TPR10, for a 10 cm x 10 cm field at depths of 20 and 10 cm for a constant SCD is used as a quantitative measure of beam quality in the Task Group 21 protocol. The RPC has measured both TPR20/TPR10 and the corresponding ratio of percent depth dose (D20/D10) at a constant SSD for 685 photon beams (4-25 MV) for most makes and models if accelerators. A strong correlation between TPR20/TPR10 and D20/D10 is presented which allows the determination of the TPR ratio from the measurement of the ratio of percent depth doses. An analysis of the uncertainty introduced in the TG-21 factors (L/rho, Pwall, Prepl) caused by the spread in the measured data and translated into the determination of the TPR ratio results in an insignificant error (< 0.3%). This empirical relationship provides an alternate technique for quantifying the beam quality defined in the TG-21 protocol without surrendering any loss of precision in output calibration. This technique may be found by those who calibrate at a fixed SSD to be an easier and quicker method.

Entities:  

Mesh:

Year:  1998        PMID: 9682206     DOI: 10.1118/1.598396

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Validation of PRIMO Monte Carlo Model of Clinac®iX 6MV Photon Beam.

Authors:  B Sarin; B Bindhu; B Saju; Raguram K Nair
Journal:  J Med Phys       Date:  2020-03-13

2.  Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission.

Authors:  Samir Didi; Abdelilah Moussa; Tayalati Yahya; Zerfaoui Mustafa
Journal:  J Med Phys       Date:  2015 Jul-Sep

3.  Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators.

Authors:  Gloria P Beyer
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

4.  Comparison of IPSM 1990 photon dosimetry code of practice with IAEA TRS-398 and AAPM TG-51.

Authors:  Silvia Vargas Castrillón; Francisco Cutanda Henríquez
Journal:  J Appl Clin Med Phys       Date:  2009-01-14       Impact factor: 2.102

5.  Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator.

Authors:  Bibekananda Mishra; Subhalaxmi Mishra; T Palani Selvam; S T Chavan; S N Pethe
Journal:  J Med Phys       Date:  2018 Jul-Sep
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.