Literature DB >> 967924

Saturation corrections for plane-parallel ionization chambers.

J Böhm.   

Abstract

Experiments described in the literature lead to different formulae for saturation corrections in ionization chambers. To elucidate the differences, saturation curves of an extrapolation chamber irradiated with beta particles from 90Sr+90Y sources have been studied experimentally. The results could be described by one formula which was a combination of known formulae for charge collection losses due to volume recombination, initial recombination and diffusion.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 967924     DOI: 10.1088/0031-9155/21/5/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Calibration of Beta-Particle Ophthalmic Applicators at the National Bureau of Standards.

Authors:  J S Pruitt
Journal:  J Res Natl Bur Stand (1977)       Date:  1986 Jul-Aug

2.  Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation.

Authors:  Ahmed M Maghraby
Journal:  Pol J Radiol       Date:  2015-11-25
  2 in total