Literature DB >> 9672016

Subdural grid implantation for intracranial EEG recording: CT and MR appearance.

M A Silberbusch1, M I Rothman, G K Bergey, G H Zoarski, M T Zagardo.   

Abstract

PURPOSE: Subdural grid arrays are used when seizure activity cannot be located by ictal scalp recordings and when functional cortical mapping is required before surgery. This study was performed to determine and compare the CT and MR imaging appearance of subdural EEG grids and to identify the types and frequency of associated complications.
METHODS: We retrospectively reviewed the medical records and imaging studies of 51 consecutive patients who underwent 54 craniotomies for subdural EEG grid implantation with either stainless steel or platinum alloy contacts between June 1988 and September 1993. Twenty-two patients had both CT and MR examinations, 27 patients had CT only, and five patients had MR imaging only. All studies were assessed for image quality and degradation by the implanted EEG grids, for intra- and extraaxial collections, and for mass effect, with differences of opinion resolved by consensus.
RESULTS: Subdural EEG grids caused extensive streak artifacts on all CT scans (corresponding directly to grid composition) and mild to moderate magnetic susceptibility artifacts on MR images. Sixteen associated complications were detected among the 54 patients imaged, including four significant extraaxial hematomas, four subfalcine or transtentorial herniations, two tension pneumocephali, two extraaxial CSF collections, two intraparenchymal hemorrhages, and one case each of cerebritis and brain abscess. In all but four cases, the detected complications were not clinically apparent and did not require specific treatment. There were no residual sequelae.
CONCLUSION: Because of extensive streak artifacts, CT showed only gross complications, such as herniation and grid displacement by extraaxial collections. MR imaging artifacts were more localized, allowing superior evaluation of subdural EEG grid placement and associated complications.

Entities:  

Mesh:

Year:  1998        PMID: 9672016      PMCID: PMC8338645     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  13 in total

1.  Rapid and fully automated visualization of subdural electrodes in the presurgical evaluation of epilepsy patients.

Authors:  Dimitri Kovalev; Joachim Spreer; Jürgen Honegger; Josef Zentner; Andreas Schulze-Bonhage; Hans-Jürgen Huppertz
Journal:  AJNR Am J Neuroradiol       Date:  2005-05       Impact factor: 3.825

2.  The utility of conductive plastic electrodes in prolonged ICU EEG monitoring.

Authors:  Rohit R Das; Brendan P Lucey; Sherry H-Y Chou; Patricio S Espinosa; Amir A Zamani; Barbara A Dworetzky; Edward B Bromfield; Jong Woo Lee
Journal:  Neurocrit Care       Date:  2008-09-20       Impact factor: 3.210

3.  Three-dimensional visualization of subdural electrodes for presurgical planning.

Authors:  Peter S LaViolette; Scott D Rand; Manoj Raghavan; Benjamin M Ellingson; Kathleen M Schmainda; Wade Mueller
Journal:  Neurosurgery       Date:  2011-03       Impact factor: 4.654

Review 4.  Placement of subdural grids in pediatric patients: technique and results.

Authors:  William E Bingaman; Juan Bulacio
Journal:  Childs Nerv Syst       Date:  2014-10-09       Impact factor: 1.475

5.  An open-source automated platform for three-dimensional visualization of subdural electrodes using CT-MRI coregistration.

Authors:  Allan A Azarion; Jue Wu; Allison Pearce; Veena T Krish; Joost Wagenaar; Weixuan Chen; Yuanjie Zheng; Hongzhi Wang; Timothy H Lucas; Brian Litt; James C Gee; Kathryn A Davis
Journal:  Epilepsia       Date:  2014-11-06       Impact factor: 5.864

6.  3D visualization of subdural electrode shift as measured at craniotomy reopening.

Authors:  Peter S LaViolette; Scott D Rand; Benjamin M Ellingson; Manoj Raghavan; Sean M Lew; Kathleen M Schmainda; Wade Mueller
Journal:  Epilepsy Res       Date:  2011-02-18       Impact factor: 3.045

7.  An Electrocorticography Grid with Conductive Nanoparticles in a Polymer Thick Film on an Organic Substrate Improves CT and MR Imaging.

Authors:  Emad Ahmadi; Husam A Katnani; Laleh Daftari Besheli; Qiang Gu; Reza Atefi; Martin Y Villeneuve; Emad Eskandar; Michael H Lev; Alexandra J Golby; Rajiv Gupta; Giorgio Bonmassar
Journal:  Radiology       Date:  2016-02-03       Impact factor: 11.105

8.  Long-term treatment with responsive brain stimulation in adults with refractory partial seizures.

Authors:  Gregory K Bergey; Martha J Morrell; Eli M Mizrahi; Alica Goldman; David King-Stephens; Dileep Nair; Shraddha Srinivasan; Barbara Jobst; Robert E Gross; Donald C Shields; Gregory Barkley; Vicenta Salanova; Piotr Olejniczak; Andrew Cole; Sydney S Cash; Katherine Noe; Robert Wharen; Gregory Worrell; Anthony M Murro; Jonathan Edwards; Michael Duchowny; David Spencer; Michael Smith; Eric Geller; Ryder Gwinn; Christopher Skidmore; Stephan Eisenschenk; Michel Berg; Christianne Heck; Paul Van Ness; Nathan Fountain; Paul Rutecki; Andrew Massey; Cormac O'Donovan; Douglas Labar; Robert B Duckrow; Lawrence J Hirsch; Tracy Courtney; Felice T Sun; Cairn G Seale
Journal:  Neurology       Date:  2015-01-23       Impact factor: 9.910

9.  PTFOS: flexible and absorbable intracranial electrodes for magnetic resonance imaging.

Authors:  Giorgio Bonmassar; Kyoko Fujimoto; Alexandra J Golby
Journal:  PLoS One       Date:  2012-09-12       Impact factor: 3.240

10.  Two-year seizure reduction in adults with medically intractable partial onset epilepsy treated with responsive neurostimulation: final results of the RNS System Pivotal trial.

Authors:  Christianne N Heck; David King-Stephens; Andrew D Massey; Dileep R Nair; Barbara C Jobst; Gregory L Barkley; Vicenta Salanova; Andrew J Cole; Michael C Smith; Ryder P Gwinn; Christopher Skidmore; Paul C Van Ness; Gregory K Bergey; Yong D Park; Ian Miller; Eric Geller; Paul A Rutecki; Richard Zimmerman; David C Spencer; Alica Goldman; Jonathan C Edwards; James W Leiphart; Robert E Wharen; James Fessler; Nathan B Fountain; Gregory A Worrell; Robert E Gross; Stephan Eisenschenk; Robert B Duckrow; Lawrence J Hirsch; Carl Bazil; Cormac A O'Donovan; Felice T Sun; Tracy A Courtney; Cairn G Seale; Martha J Morrell
Journal:  Epilepsia       Date:  2014-02-22       Impact factor: 5.864

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.