Unreliability of the infrared tympanic thermometer in clinical practice: a comparative study with oral mercury and oral electronic thermometers.

J G Modell1, C R Katholi, S M Kumaramangalam, E C Hudson, D Graham.   

Abstract

BACKGROUND: This study was designed to determine the magnitude and frequency of measurement errors with infrared tympanic thermometers in the clinical setting.
METHODS: In a convenience sample of 137 adult inpatients, we compared body temperatures measured by a Diatek 9000 Infrared Aural Thermometer and an IVAC 2090 CoreCheck Tympanic Thermometer between themselves, in right versus left ears, and against concurrently measured oral temperatures using both an electronic thermoprobe and conventional glass mercury thermometer.
RESULTS: There was a significant between-brand difference of 0.6 degrees C (IVAC <Diatek) for both ears; right-versus-left ear temperature group differences were not statistically significant with either thermometer, but ear-to-ear temperature variation was large, with individual right-left differences as great as 2.5 degrees C; significant differences were found between oral and ear temperatures with the IVAC thermometer (0.5 degrees C, ear <oral) but not with the Diatek thermometer; no significant difference was found between oral electronic and mercury temperature measurements; and cerumen had no significant effect on ear temperature measurements.
CONCLUSIONS: The variability and inaccuracy of temperatures measured by the infrared tympanic thermometers were sufficiently large to suggest that the use of these devices for routine thermometry may be potentially hazardous.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9671836     DOI: 10.1097/00007611-199807000-00008

Source DB:  PubMed          Journal:  South Med J        ISSN: 0038-4348            Impact factor:   0.954


  6 in total

1.  Effect of pre-warmed intravenous fluids on perioperative hypothermia and shivering after ambulatory surgery under monitored anesthesia care.

Authors:  Gahyun Kim; Myung Hee Kim; Sangmin M Lee; Soo Joo Choi; Young Hee Shin; Hee Joon Jeong
Journal:  J Anesth       Date:  2014-04-01       Impact factor: 2.078

2.  Temperature measurement in paediatrics.

Authors: 
Journal:  Paediatr Child Health       Date:  2000-07       Impact factor: 2.253

3.  Effect of the status after ear surgery and ear pathology on the results of infrared tympanic thermometry.

Authors:  Frank Schmäl; Marjolein Loh-van den Brink; Wolfgang Stoll
Journal:  Eur Arch Otorhinolaryngol       Date:  2005-07-06       Impact factor: 2.503

4.  Comparability of tympanic and oral mercury thermometers at high ambient temperatures.

Authors:  Amy L Chue; Rachael L Moore; Andrew Cavey; Elizabeth A Ashley; Kasia Stepniewska; François Nosten; Rose McGready
Journal:  BMC Res Notes       Date:  2012-07-16

5.  Effect of pre-warming on perioperative hypothermia during holmium laser enucleation of the prostate under spinal anesthesia: a prospective randomized controlled trial.

Authors:  Joo-Hyun Jun; Mi Hwa Chung; Eun Mi Kim; In-Jung Jun; Jung Hwa Kim; Joon-Sang Hyeon; Mi Hyeon Lee; Hye Sun Lee; Eun Mi Choi
Journal:  BMC Anesthesiol       Date:  2018-12-22       Impact factor: 2.217

6.  Effectiveness of Combined Strategies for the Prevention of Hypothermia Measured by Noninvasive Zero-Heat Flux Thermometer During Cesarean Section.

Authors:  Antonella Cotoia; Paola Sara Mariotti; Claudia Ferialdi; Pasquale Del Vecchio; Renata Beck; Simona Zaami; Gilda Cinnella
Journal:  Front Med (Lausanne)       Date:  2021-12-23
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.