Calculation of absorbed dose distributions from dynamic wedges.

P Storchi1, E Woudstra, P Verlinde, K A Johansson, A Samuelsson.   

Abstract

In radiotherapy with photon beams, the use of dynamic wedges, which are obtained by the movement of one of the jaws, offers an increasing flexibility relative to the traditional use of metal wedges. But it is a disadvantage for the measurement of absorbed dose distributions, because the absorbed dose at each measurement point can only be obtained after a complete movement of the jaw. Consequently, for radiotherapy planning, an algorithm should be available that does not require measurements for any specific dynamically wedged beam, but is based on only a modest number of measurements. In this paper, an algorithm for the calculation of the dose distribution from dynamic wedges is described. This algorithm uses the convolution of pencil beam kernels with a non-uniform field function. These pencil beam kernels are derived from empirical data resulting from measurements of the open beam only.

Mesh:

Year:  1998        PMID: 9651020     DOI: 10.1088/0031-9155/43/6/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Comparison between measured and calculated dynamic wedge dose distributions using the anisotropic analytic algorithm and pencil-beam convolution.

Authors:  Paola Francisca Caprile; Carlos Daniel Venencia; Pelayo Besa
Journal:  J Appl Clin Med Phys       Date:  2006-08-08       Impact factor: 2.102