On-line assessment of delivered dialysis dose.

S Di Filippo1, S Andrulli, C Manzoni, M Corti, F Locatelli.   

Abstract

BACKGROUND: The adequacy of the delivered dialysis dose is essential to prevent patient morbidity and mortality. The determination of effective ionic dialysance (D) is easy, non-invasive and inexpensive, and its use instead of effective urea clearance (K) in kinetically determining apparent" urea distribution volume (Vt) is likely to lead to a correct Kt/V, even though the Vt value may be incorrect. The aim of this study was to test the possibility of using the measurement of D to monitor Kt/V on-line during each dialysis treatment.
METHODS: Forty-four patients were dialyzed using a monitor equipped with specially designed "Diascan Module" (COT; Hospal) that measures effective D by means of a single conductivity probe. Vt was calculated according to the SPVV three BUN method urea kinetic model using D instead of K values. One month later, Kt/V was calculated as Dt/V, using actual D and T values and the predetermined Vt values updated for the current final body wt. Both the Dt/V and Kt/V determined according to the Smye and Daugirdas methods were compared with the Kt/V determined using the SPVV kinetic model (Kt/Veq)
RESULTS: The Kt/V values calculated using ionic dialysance and predetermined Vt were approximately equivalent to those of Kt/Veq (1.14 +/- 0.16 vs. 1.14 +/- 0.17, mean difference 0.00 +/- 0.07), as were those determined according to the Smye and Daugirdas methods (1.10 +/- 0.18 and 1.13 +/- 0.17, mean difference -0.03 +/- 0.06 and 0.01 +/- 0.06, respectively).
CONCLUSION: Once Vt has been determined, the evaluation of ionic dialysance in stable patients makes it possible to calculate the Kt/V accurately at each dialysis session without blood or dialysate sampling, and at no additional cost.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9648087     DOI: 10.1046/j.1523-1755.1998.00979.x

Source DB:  PubMed          Journal:  Kidney Int        ISSN: 0085-2538            Impact factor:   10.612


  2 in total

1.  Interventions to improve hemodialysis adequacy: protocols based on real-time monitoring of dialysate solute clearance.

Authors:  Edward A Ross; Jennifer L Paugh-Miller; Robert W Nappo
Journal:  Clin Kidney J       Date:  2017-10-25

2.  Correlation between Dt/V derived from ionic dialysance and blood-driven Kt/V of urea in African-American hemodialysis patients, based on body weight and ultrafiltration volume.

Authors:  Wihib Gebregeorgis; Zeenat Yousuf Bhat; Nishigandha Pradhan; Stephen D Migdal; Lakshminarayanan Nandagopal; Reddy Singasani; Tehmina Mushtaq; Ronald Thomas; Yahya M Osman Malik
Journal:  Clin Kidney J       Date:  2018-01-31
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.