| Literature DB >> 9639536 |
J L Dage1, B L Ackermann, H B Halsall.
Abstract
A simple, fast and sensitive method was developed to verify the presence of the sialyl Lewis(x) antigen on an N-linked glycoprotein. High performance liquid chromatography-electrospray mass spectrometry (HPLC-ESI/MS) was used to identify which of the five N-linked glycosylation sites of human plasma alpha1-acid-glycoprotein (orosomucoid, OMD) contain the sialyl Lewis(x) antigen. OMD was digested with proteolytic enzymes and analyzed by reversed phase chromatography coupled with on-line ESI/MS. A tandem mass spectrometry experiment was designed to detect the presence of the sialyl Lewis(x) antigen based on the observation of an 803 mass to charge ratio ( m/z ) ion produced in the intermediate pressure region of the ESI interface. The ESI/MS signal at m/z 803 is consistent with an oxonium ion for a glycan structure containing NeuAc, Gal, GlcNAc, and Fuc. The identity of the m/z 803 ion was confirmed by ESI/MS/MS analysis of the m/z 803 fragment ion and comparison with a sialyl Lewis(x) standard. The stereochemistry and linkage positions were assigned using previous NMR analysis but could be determined with permethylation analysis if necessary. The analysis of OMD gave a pattern showing signal for the sialyl Lewis(x) antigen coeluting with each of the five N-linked glycopeptides. The ability to monitor sialyl Lewis(x) expression at each of the five sites is of interest in the study of OMD's role in inflammatory diseases.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9639536 DOI: 10.1093/glycob/8.8.755
Source DB: PubMed Journal: Glycobiology ISSN: 0959-6658 Impact factor: 4.313